FDA Enforcement Class II Ongoing

REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.

Recall: Z-2079-2025 · Reported July 9, 2025

Enforcement

Recall Number
Z-2079-2025
Event ID
96922
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton, Dickinson and Company, BD Bio Sciences
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 9, 2025
Initiation Date
June 3, 2025
Classification Date
July 3, 2025
Address
155 N Mccarthy Blvd, Milpitas, CA, 95035-5102, United States

Description

REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.

Reason

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

Code Info

Lot # 85637 & 85638/ UDI: 382906662897

Distribution

US: Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Hong Kong, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Peru, Pakistan, Singapore, El Salvador, Taiwan OUS:

Quantity

278 units