Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128
Enforcement
- Recall Number
- Z-2079-2012
- Event ID
- 61974
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Getinge USA Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 1, 2012
- Initiation Date
- May 25, 2012
- Classification Date
- July 26, 2012
- Termination Date
- May 18, 2016
- Address
- 1777 E Henrietta Rd, Rochester, NY, 14623-3133, United States
Description
Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128
Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error.
Serial numbers: W50024258; W50024259; W50026573; W50025790; W50017381; W50020296; W50028138; W50020952
Worldwide Distribution - USA including Maryland, Indiana, New Mexico, Massachuesetts, Texas, Kansas and the country of Canada
8 washers