U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions

Recall: Z-2076-2019 · Reported July 31, 2019

Enforcement

Recall Number
Z-2076-2019
Event ID
83310
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
LABORIE MEDICAL TECHNOLOGIES, CORP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2019
Initiation Date
July 17, 2019
Classification Date
July 25, 2019
Termination Date
July 10, 2023
Address
400 Avenue D Ste 10, N/A, Williston, VT, 05495-7828, United States

Description

U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions

Reason

Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard

Code Info

Serial Numbers: PE18-7SBL1452, PE18-7SBL 1414

Distribution

US distribution in AZ, and PA.

Quantity

2 US