FDA Enforcement
Class II
Terminated
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Recall: Z-2070-2014
·
Reported July 23, 2014
Enforcement
- Recall Number
- Z-2070-2014
- Event ID
- 68573
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Excelsior Medical Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 23, 2014
- Initiation Date
- June 5, 2014
- Classification Date
- July 16, 2014
- Termination Date
- August 19, 2015
- Address
- 1933 Heck Ave, N/A, Neptune, NJ, 07753, United States
Description
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Reason
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
Code Info
Product Code PA03 - Lot Nos.: 03030813, 03050813 and 03091213 Product Code BX03, Lot Nos.: 64031513, 64073113, 64081313, 64090513
Distribution
Worldwide Distribution - USA (nationwide) and Canada.
Quantity
4250 tubing sets (PA03 - 78 cases; BX03 - 92 cases (each case contains 25 tubing sets))