DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 1402 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
Enforcement
- Recall Number
- Z-2067-2020
- Event ID
- 85485
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Merit Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 27, 2020
- Initiation Date
- February 13, 2020
- Classification Date
- May 21, 2020
- Termination Date
- February 26, 2021
- Address
- 1600 W Merit Pkwy, N/A, South Jordan, UT, 84095-2416, United States
Description
DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 1402 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).
Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the procedure to be rescheduled.
Lot number: AAD-1801-16 and UDI: (01)00884450374399(17)221231(10)AAD-1801-16; Lot number: AAD-1801-17 and UDI: (01)00884450374399(17)221231(10)AAD-1801-17; Lot number: AAD-1803-01 and UDI: (01)00884450374399(17)230331(10)AAD-1803-01; Lot number: AAD-1804-05 and UDI: (01)00884450374399(17)230430(10)AAD-1804-05; Lot number: AAD-1901-07 and UDI: (01)00884450374399(17)231231(10)AAD-1901-07; Lot number: AAD-1904-04 and UDI: (01)00884450374399(17)240430(10)AAD-1904-04.
Worldwide distribution: US Nationwide distribution including in the states of (TX, UT, PA, KY, TX, AR, CA, and PA) which included one (1) government/military consignee in KY; and the countries of (Germany, Switzerland, and New Zealand).
228