Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the subject of this recall, and the recalled BD Alaris Pump Infusion Set (RES#82951) Product Usage: The Vyaire enFlow Disposable Cartridges contain a heating element that is used to help raise or maintain a patient's body temperature.
Enforcement
- Recall Number
- Z-2066-2019
- Event ID
- 82940
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medline Industries Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 21, 2019
- Initiation Date
- May 3, 2019
- Classification Date
- August 12, 2019
- Termination Date
- March 1, 2021
- Address
- Three Lakes Drive, N/A, Northfield, IL, 60093, United States
Description
Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the subject of this recall, and the recalled BD Alaris Pump Infusion Set (RES#82951) Product Usage: The Vyaire enFlow Disposable Cartridges contain a heating element that is used to help raise or maintain a patient's body temperature.
The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.
Lots 2018090690 2018101190 2018101590 2018121190 2019013190 2019030490 2019040190
One distributor located in Illinois.
4,680 kits