FDA Enforcement Class II Ongoing

ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;

Recall: Z-2063-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-2063-2026
Event ID
98674
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Erbe Medical, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
May 13, 2026
Initiation Date
March 31, 2026
Classification Date
May 1, 2026
Address
405 W Geneva Dr, Tempe, AZ, 85282-2003, United States

Description

ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;

Reason

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

Code Info

REF/UDI-DI/Lots: 20325-201/04065655000603/W4466253, W4466254, W4466256, W4466257, W4466258, WO471080, WO471081, WO471083, WO471084, WO473975, WO473976, WO477586, WO477587, WO477588, WO477589, WO477590; 20325-203/04065655001211/WO473986

Distribution

US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.

Quantity

23,359