BETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number: 110045160
Enforcement
- Recall Number
- Z-2061-2023
- Event ID
- 92410
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 12, 2023
- Initiation Date
- May 4, 2023
- Classification Date
- July 3, 2023
- Address
- Kibbutz Gatton, N/A, GAATON, N/A, N/A, Israel
Description
BETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number: 110045160
Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.
UDI-DI: 10818674025772 Lot Numbers: 21E28, 21R15, 21R16, 21R32, 21R35, 22C19, 22E29, 22F10, 22F35, 22J07, 22J08, 22J09, 22J18, 22J19, 22J27, 22J28, 22J33, 22K23, 22K24, 22K32, 22K33, 22K38, 22M03, 22M04, 22M24, 22P03, 22P09, 22P10, 22P11, 22P13, 22P14, 22P16
Worldwide distribution - US Nationwide and the countries of Australia, Japan, Netherlands.