FDA Enforcement Class II Ongoing

BETTA LINK SR KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number:110045154

Recall: Z-2060-2023 · Reported July 12, 2023

Enforcement

Recall Number
Z-2060-2023
Event ID
92410
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 12, 2023
Initiation Date
May 4, 2023
Classification Date
July 3, 2023
Address
Kibbutz Gatton, N/A, GAATON, N/A, N/A, Israel

Description

BETTA LINK SR KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number:110045154

Reason

Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.

Code Info

UDI-DI: 0818674025765 Lot Numbers: 21R01, 21R02, 21R03, 22C02, 22C03, 22C06, 22C08, 22E01, 22E02, 22F01, 22F02, 22F03, 22J01, 22J02, 22J03, 22K01, 22K02, 22K03, 22K04, 22K05, 22K07, 22K08, 22K09, 22K10, 22K11, 22M01, 22M02, 22M03, 22M04, 22N01, 22P01, 22P02, 22P03, 22P04, 22P05, 22R01

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Japan, Netherlands.

Quantity

N/A