FDA Enforcement
Class II
Terminated
Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F
Recall: Z-2059-2019
·
Reported July 31, 2019
Enforcement
- Recall Number
- Z-2059-2019
- Event ID
- 83026
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zyno Medical LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 31, 2019
- Initiation Date
- May 29, 2019
- Classification Date
- July 23, 2019
- Termination Date
- June 22, 2020
- Address
- 177 Pine St, N/A, Natick, MA, 01760-1331, United States
Description
Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F
Reason
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
Code Info
Lot Numbers: 18046309, 18026344 UDl/UPC/GTIN: 00814371020204
Distribution
MA
Quantity
360 cases (50/case)=18,000