FDA Enforcement Class II Terminated

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Recall: Z-2058-2019 · Reported July 31, 2019

Enforcement

Recall Number
Z-2058-2019
Event ID
83026
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zyno Medical LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 31, 2019
Initiation Date
May 29, 2019
Classification Date
July 23, 2019
Termination Date
June 22, 2020
Address
177 Pine St, N/A, Natick, MA, 01760-1331, United States

Description

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Reason

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Code Info

Lot Numbers:18037128,18065375, 19015969 UDl/UPC/GTIN: 00814371020198

Distribution

MA

Quantity

354 cases (50/case)=17700