FDA Enforcement
Class II
Terminated
Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research.
Recall: Z-2058-2018
·
Reported June 20, 2018
Enforcement
- Recall Number
- Z-2058-2018
- Event ID
- 80115
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zeiss, Carl Inc
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- June 20, 2018
- Initiation Date
- May 8, 2018
- Classification Date
- July 11, 2018
- Termination Date
- April 21, 2020
- Address
- 1 Zeiss Dr, N/A, Thornwood, NY, 10594-1939, United States
Description
Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research.
Reason
In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.
Code Info
Model No. Cell Observer SD, Laser TIRF, and DirectFRAP Laser Scanning Microscopes
Distribution
USA (nationwide)
Quantity
1070 units to US