FDA Enforcement Class II Terminated

Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.

Recall: Z-2057-2014 · Reported July 23, 2014

Enforcement

Recall Number
Z-2057-2014
Event ID
68583
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Remel Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 23, 2014
Initiation Date
June 11, 2014
Classification Date
July 15, 2014
Termination Date
August 28, 2014
Address
12076 Santa Fe Trail Dr, N/A, Lenexa, KS, 66215-3519, United States

Description

Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.

Reason

Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae.

Code Info

Segments 1, 5, and 6 only of Lot 492571, Exp. 2015-10

Distribution

Distribution was made to CO, IL, IN, NE, NJ, PA, SD, and a government facility located in VA.

Quantity

76/50-bottle boxes