FDA Enforcement Class I Ongoing

Heartware HVAD Pump Implant Kit, Product (REF) Number 1104

Recall: Z-2056-2020 · Reported June 3, 2020

Enforcement

Recall Number
Z-2056-2020
Event ID
85397
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Heartware, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 3, 2020
Initiation Date
April 3, 2020
Classification Date
May 22, 2020
Address
14400 Nw 60th Ave, N/A, Miami Lakes, FL, 33014-2807, United States

Description

Heartware HVAD Pump Implant Kit, Product (REF) Number 1104

Reason

The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.

Code Info

GTIN 00888707002028, Serial Numbers: HW28848, hw29011, HW29154, HW29211, HW29248, HW29249, HW29273, HW29329, HW29426, HW29465, HW29472, HW29522, HW29534, HW29548, HW29549, HW29567, HW29588, HW29593, HW29602, HW29612, HW29667, HW29700, HW29786, HW29815, HW29915, HW29928, HW29949, HW30039, HW30082, HW30083, HW30150, HW30153, HW30208, HW30211, HW30226, HW30320, HW30342, HW30358, HW30360, HW30363, HW30378, HW30380, HW30382, HW30390, HW30398, HW30400, HW30402, HW30406, HW30410, HW30430, HW30475, HW30484, HW30494, HW30498, HW30595, HW30596, HW30597, HW30601, HW30604, HW30651, HW30768, HW30769, HW30770, HW30771, HW30814, HW30821, HW30829, HW30863, HW30865, HW30867, HW30871, HW30953, HW30961, HW30962, HW30966, HW30971, HW30972, HW30974, HW31020, HW31076, HW31101, HW31103, HW31104, HW31107, HW31113, HW31115, HW31117, HW31119, HW31128, HW31132, HW31134, HW31192, HW31219, HW31223, HW31232, HW31235, HW31242, HW31244, HW31249, HW31254, HW31255, HW31260, HW31261, HW31262, HW31267, HW31383, HW31387, HW31388, HW31399, HW31409, HW31410, HW31411, HW31412, HW31414, hw31445, HW31446, hw31447, hw31448, hw31449, hw31451, HW31452, hw31453, hw31454, HW31455, HW31456, HW31457, HW31458, HW31541, HW31543, HW31544, HW31545, HW31547, HW31548, HW31549, HW31550, HW31552, HW31559, HW31564, HW31571, HW31573, HW31578, HW31579, HW31686, HW31694, HW31697, HW31731, HW31749, HW31750, HW31777, HW31778, HW31810, HW31819, HW31830, HW31831, HW31832, HW31834, HW31835, HW31836, HW31838, HW31839, HW31840, HW31842, HW31843, HW31844, HW31845, HW31846, HW31847, HW31849, HW31855, HW31858, HW31876, HW31883, HW31944, HW31947, HW31979, HW31981, HW31986, HW31987, HW31988, HW31989, HW31991, HW32068, HW32116, HW32121, HW32141, HW32143, HW32156, HW32159, HW32161, HW32162, HW32164, HW32167, HW32172, HW32179, HW32189, HW32190, HW32192, HW32193, HW32194, HW32197, HW32199, HW32201, HW32204, HW32206, HW32209, HW32210, HW32214, HW32215, HW32216, HW32218, HW32220, HW32222, HW32223, HW32224, HW32228, HW32229, HW32230, HW32231, HW32248, HW32258, HW32259, HW32260, HW32261, HW32273, HW32279, HW32281, HW32285, HW32286, HW32290, HW32292, HW32315, HW32335, HW32357, HW32386, HW32393, HW32400, HW32419, HW32423, HW32426, HW32427, HW32457, HW32458, HW32459, HW32464, HW32465, HW32468, HW32470, HW32471, HW32472, HW32474, HW32479, HW32499, HW32504, HW32508, HW32509, HW32510, HW32511, HW32516, HW32534, HW34001, HW34002, HW34025, HW34034, HW34035, HW34036, HW34037, HW34038, HW34040, HW34041, HW34042, HW34043, HW34044, HW34045, HW34046, HW34047, HW34048, HW34051, HW34066, HW34096, HW34100, HW34104, HW34106, HW34109, HW34110, HW34111, HW34113, HW34117, HW34121, HW34126, HW34151, HW34169, HW34179, HW34181, HW34186, HW34187, HW34192, HW34193, HW34196, HW34197, HW34202, HW34204, HW34205, HW34207, HW34209, HW34211, HW34226, HW34317, HW34319, HW34326, HW34328, HW34329, HW34330, HW34331, HW34336, HW34342, HW34344, HW34347, HW34368, HW34369, HW34371, HW34372, HW34373, HW34375, HW34376, HW34377, HW34378, HW34379, HW34425, HW34427, HW34428, HW34429, HW34432, HW34446, HW34495, HW34519, HW34520, HW34521, HW34522, HW34523, HW34546, HW34547, HW34550, HW34551, HW34555, HW34556, HW34588, HW34590, HW34596, HW34603, HW34605, HW34609, HW34610, HW34643, HW34673, HW34752, HW34768, HW34805, HW34806, HW34808, HW34810, HW34824, HW34840, HW34842, HW34845, HW34866, HW34868, HW34877, HW34902, HW34903, HW34904, HW34908, HW34910, HW34913, HW34914, HW34923, HW34924, HW34925, HW34928, HW34933, HW34934, HW34938, HW34939, HW34944, HW34946, HW34947, HW34948, HW34949, HW34957, HW34979, HW34990, HW34997, HW35012, HW35032, HW35099, HW35122, HW35123, HW35124, HW35125, HW35127, HW35128, HW35131, HW35133, HW35134, HW35136, HW35138, HW35141, HW35142, HW35143, HW35161, HW35189, HW35191, HW35204, HW35205, HW35209, HW35211, HW35212, HW35213, HW35214, HW35215, HW35220, HW35284, HW35305, HW35317, HW35319, HW35321, HW35324, HW35325, HW35326, HW35332, HW35334, HW35337, HW35339, HW35340, HW35341, HW35357, HW35360, HW35361, HW35367, HW35385, HW35464, HW35477, HW35478, HW35479, HW35480, HW35481, HW35482, HW35483, HW35487, HW35488, HW35489, HW35492, HW35493, HW35494, HW35495, HW35499, HW35500, HW35501, HW35502, HW35504, HW35506, HW35508, HW35534, HW35535, HW35539, HW35540, HW35541, HW35542, HW35543, HW35545, HW35548, HW35563, HW35577, HW35578, HW35580, HW35582, HW35583, HW35584, HW35586, HW35587, HW35588, HW35624, HW35625, HW35626, HW35627, HW35628, HW35629, HW35630, HW35631, HW35632, HW35633, HW35634, HW35635, HW35636, HW35637, HW35639, HW35641, HW35644, HW35646, HW35647, HW35668, HW35669, HW35670, HW35671, HW35673, HW35674, HW35676, HW35677, HW35678, HW35680, HW35681, HW35682, HW35683, HW35684, HW35686, HW35687, HW35688, HW35689, HW35690, HW35691, HW35692, HW35693, HW35695, HW35696, HW35697, HW35698, HW35701, HW35702, HW35703, HW35704, HW35705, HW35707, HW35708, HW35730, HW35731 GTIN 00888707002387, Serial Numbers: HW28851, HW29783, HW29793, HW29794, HW29827, HW29870, HW29918, HW30045, HW30046, HW30505, HW30765, HW30842, HW30844, HW30845, HW30846, hw30978, HW31075, HW31077, HW31080, hw31089, hw31091, HW31095, HW31161, HW31342, HW31348, HW31555, HW31556, HW31557, HW31683, HW31860, HW31903, HW31920, HW31924, HW31926, HW32026, HW32027, HW32028, HW32149, HW32153, HW32163, HW32282, HW32513, HW32514, HW32562, HW32563, HW34004, HW34006, HW34091, HW34092, HW34125, HW34272, HW34273, HW34274, HW34349, HW34350, HW34351, HW34352, HW34357, HW34358, HW34359, HW34366, HW34374, HW34436, HW34953, HW34968, HW35080, HW35144, HW35145, HW35148, HW35327, HW35342, HW35344, HW35537, HW35596, HW35603, HW35604, HW35605, HW35606, HW35607, HW35713, HW35714, HW35717, HW35718, HW35721, HW35723 GTIN 00888707001649, Serial Numbers: HW29132, HW29232, hw29653, HW29682, HW29684, HW29705, HW29864, HW29985, HW29991, HW30096, HW30100, HW30145, HW30146, HW30373, HW30374, HW30421, HW30525, HW30529, HW30726, HW30738, HW30836, HW30840, HW30862, HW30939, hw30949, hw30950, HW31048, HW31071, HW31082, HW31088, hw31133, hw31394, HW31415, HW31417, HW31420, HW31421, HW31423, hw31483, hw31505, hw31506, hw31507, hw31510, HW31511, hw31512, hw31513, hw31515, hw31516, hw31517, HW31546, HW31551, HW31554, HW31566, HW31587, HW31588, HW31589, HW31590, HW31592, HW31632, HW31639, HW31685, HW31795, HW31801, HW31806, HW31808, HW31809, HW31811, HW31812, HW31813, HW31815, HW31891, HW31892, HW31893, HW31919, HW31921, HW31973, HW31982, HW31985, HW32000, HW32001, HW32014, HW32015, HW32018, HW32021, HW32022, HW32024, HW32042, HW32050, HW32055, HW32056, HW32064, HW32081, HW32094, HW32112, HW32123, HW32138, HW32147, HW32186, HW32245, HW32255, HW32264, HW32265, HW32268, HW32270, HW32301, HW32394, HW32410, HW32420, HW32558, HW32559, HW32560, HW32561, HW34029, HW34031, HW34093, HW34094, HW34132, HW34148, HW34180, HW34266, HW34268, HW34269, HW34270, HW34271, HW34296, HW34301, HW34324, HW34327, HW34340, HW34341, HW34343, HW34345, HW34346, HW34348, HW34418, HW34420, HW34431, HW34507, HW34508, HW34641, HW34692, HW34745, HW34764, HW34836, HW34906, HW34932, HW34936, HW34940, HW34941, HW34963, HW34977, HW34992, HW34994, HW35019, HW35021, HW35033, HW35035, HW35038, HW35041, HW35065, HW35068, HW35085, HW35090, HW35094, HW35114, HW35147, HW35157, HW35196, HW35197, HW35210, HW35237, HW35250, HW35290, HW35308, HW35309, HW35310, HW35311, HW35312, HW35313, HW35314, HW35315, HW35318, HW35320, HW35328, HW35329, HW35330, HW35333, HW35338, HW35349, HW35365, HW35366, HW35368, HW35370, HW35371, HW35372, HW35373, HW35375, HW35378, HW35402, HW35404, HW35405, HW35408, HW35409, HW35411, HW35412, HW35418, HW35421, HW35423, HW35465, HW35470, HW35475, HW35476, HW35485, HW35486, HW35546, HW35547, HW35550, HW35551, HW35552, HW35553, HW35554, HW35555, HW35556, HW35557, HW35558, HW35559, HW35560, HW35564, HW35598, HW35599, HW35602, HW35640, HW35657, HW35658, HW35659, HW35662, HW35663, HW35664, HW35665, HW35666, HW35667, HW35672, HW35694, HW35719, HW35750, HW35753, HW36121, HW36154, HW36165, HW36168 GTIN 00888707000581, Serial Numbers: HW29203, HW29425, HW29427, HW29428, HW29430, HW29439, HW29517, HW29679, HW29680, HW30589, HW30638, HW30706, HW30876, HW31144, HW31187, hw31376, hw31380, HW31442, HW31727, HW31998, HW32023, HW32416, HW34017 GTIN 00888707002417, Serial Numbers: HW29478, HW30381, HW30831, HW34380, HW34421, HW35348, HW35613, HW35756, HW35761, HW35763 GTIN 00888707002493, Serial Numbers: HW29483, HW29697, HW29698, hw31178, HW31520, HW31821, HW31823, HW31827, HW32086, HW32242, HW32246, HW32518, HW32519, HW34021, HW34365, HW34367, HW34792, HW34915, HW35155, HW35158, HW35252, HW35343, HW35345, HW35346, HW35350, HW35400, HW35401, HW35529, HW35530, HW35574, HW35591, HW35660, HW35661, HW35785 GTIN 00888707002400, Serial Numbers: HW29518, HW29693, HW30092, HW32257, HW34012, HW34210, HW35034, HW35154, HW35573, HW35653 GTIN 00888707002547, Serial Numbers: HW29561, HW29601, HW29931, HW30502, HW30593, HW30619, HW30632, HW30783, HW30796, HW30819, HW30826, HW31136, HW31140, HW31141, HW31156, HW31157, HW31158, HW31382, HW31393, HW31406, HW31408, HW31413, HW31416, HW31426, HW31429, HW31553, HW31791, HW31937, HW31994, HW32118, HW32269, HW32309, HW32336, HW32347, HW32373, HW32385, HW32522, HW32523, HW32524, HW32554, HW32557, HW34003, HW34007, HW34008, HW34049, HW34050, HW34053, HW34054, HW34382, HW34383, HW34513, HW34514, HW34516, HW34517, HW34518, HW35251, HW35275 GTIN 00888707000680, Serial Numbers: HW29571, HW29578, HW29845, HW30635, HW30639, HW30711, HW31150 GTIN 00888707002394, Serial Numbers: HW30469, HW30506, HW30508, HW30509, HW30510, HW30511, HW30512, HW30655, HW31580, HW31582, HW31583, HW31584, HW31992, HW31993, HW32200, HW32207, HW32217, HW32238, HW32239, HW32240, HW32515, HW32517, HW34022, HW34023, HW34395, HW34461, HW34557, HW34561, HW34582, HW34751, HW34950, HW34955, HW34991, HW34996, HW35016, HW35149, HW35150, HW35186, HW35336, HW35386, HW35397, HW35484, HW35509, HW35510, HW35538, HW35566, HW35609, HW35610, HW35611, HW35648, HW35649, HW35651, HW35679, HW35699, HW35700, HW35706, HW35776, HW35779 GTIN 00888707000239, Serial Numbers: HW30634, HW30645, HW30877, HW34737 GTIN 00888707002530, Serial Numbers: HW30797, HW31108, HW31440, HW31828, HW31928, HW32182, HW32253, HW32263, HW34559, HW35000, HW35162, HW35532, HW35533, HW35590 GTIN 00888707002523, Serial Numbers: HW30859, HW31703, HW31704, HW31859, HW32030, HW32036, HW32040, HW32102, HW32520, HW32521, HW34018, HW34020, HW34027, HW34028, HW34986, HW35040, HW35043, HW35049, HW35216, HW35217, HW35218, HW35490, HW35531, HW35619, HW35786 GTIN 00888707002431, Serial Numbers: hw30979, hw31123, HW32029, HW32180, HW34024, HW34381, HW34615, HW35119, HW35733 GTIN 00888707002585, Serial Numbers: hw31079, hw31081, hw31168, hw31170, HW35095, HW35164, HW35351, HW35353, HW35355, HW35496, HW35497, HW35575, HW35576 GTIN 00888707002516, Serial Numbers: hw31147, hw31162, HW32507, HW34524, HW35474 GTIN 00888707006330, Serial Numbers: HW31419, HW31699, HW31733, HW32505, HW35393, HW35396, HW35726, HW35749, HW35766, HW35769, HW35784, HW35829, HW35872, HW36005, HW36007, HW36008, HW36009, HW36010, HW36013, HW36014, HW36015, HW36018, HW36019, HW36020, HW36021, HW36067, HW36068, HW36069, HW36070, HW36071, HW36131, HW36132, HW36138, HW36174, HW36175, HW36176, HW36177, HW36178, HW36179, HW36180, HW36182, HW36184, HW36185, HW36186, HW36192, HW36193, HW36194, HW36195, HW36196, HW36197, HW36198, HW36199, HW36200, HW36201, HW36224, HW36225, HW36226, HW36227, HW36228, HW36229, HW36230, HW36231, HW36232, HW36233, HW36248, HW36249, HW36251, HW36252, HW36253, HW36254, HW36255, HW36256, HW36258, HW36259, HW36260, HW36276, HW36298, HW36299, HW36300, HW36303, HW36305, HW36306, HW36307, HW36309, HW36310, HW36313, HW36314, HW36316, HW36317, HW36318, HW36319, HW36320, HW36321, HW36323, HW36324, HW36325, HW36326, HW36328, HW36332, HW36333, HW36334, HW36335, HW36337, HW36354, HW36355, HW36356, HW36357, HW36358, HW36359, HW36360, HW36361, HW36362, HW36363, HW36396, HW36397, HW36398, HW36399, HW36400, HW36401, HW36403, HW36404, HW36405, HW36406, HW36407, HW36408, HW36409, HW36410, HW36411, HW36412, HW36419, HW36420, HW36428, HW36429, HW36430, HW36432, HW36433, HW36434, HW36435, HW36436, HW36437, HW36438, HW36500, HW36501, HW36502, HW36503, HW36504, HW36505, HW36506, HW36507, HW36508, HW36509, HW36510, HW36511, HW36513, HW36514, HW36515, HW36530, HW36553, HW36554, HW36555, HW36556, HW36557, HW36558, HW36559, HW36560, HW36561, HW36562, HW36563, HW36564, HW36595, HW36596, HW36597, HW36598, HW36599, HW36600, HW36601, HW36602, HW36603, HW36624, HW36625, HW36626, HW36628, HW36630, HW36631, HW36632, HW36633, HW36636, HW36637, HW36655, HW36659, HW36660, HW36662, HW36663, HW36664, HW36669, HW36670, HW36671, HW36672, HW36703, HW36704, HW36705, HW36708, HW36709, HW36710, HW36711, HW36712, HW36713, HW36853, HW36854, HW36855, HW36898, HW36899, HW36900, HW36903, HW36919, HW36921 GTIN 00888707002486, Serial Numbers: HW31444, HW31569, HW31673, HW31814, HW31817, HW31818, HW31820, HW31824, HW31825, HW31826, HW31872, HW31873, HW31875, HW31877, HW31927, HW31932, HW31934, HW31939, HW31940, HW31941, HW31995, HW31996, HW31997, HW31999, HW32002, HW32008, HW32009, HW32010, HW32011, HW32013, HW32032, HW32034, HW32038, HW32046, HW32053, HW32057, HW32062, HW32065, HW32071, HW32085, HW32109, HW32165, HW32166, HW32195, HW32205, HW32211, HW32221, HW32236, HW32243, HW32250, HW32506, HW34015, HW34122, HW34123, HW34124, HW34251, HW34252, HW34275, HW34277, HW34280, HW34354, HW34355, HW34356, HW34361, HW34364, HW34433, HW34434, HW34437, HW34439, HW34440, HW34442, HW34443, HW34444, HW34452, HW34454, HW34619, HW34620, HW34621, HW34622, HW34623, HW34624, HW34629, HW34630, HW34631, HW34636, HW34656, HW34719, HW34738, HW34739, HW34742, HW34748, HW34749, HW34750, HW34753, HW34757, HW34758, HW34761, HW34762, HW34767, HW34769, HW34779, HW34818, HW34853, HW34867, HW34873, HW34879, HW34880, HW34882, HW34885, HW34888, HW34890, HW34897, HW34898, HW34900, HW34905, HW34909, HW34911, HW34912, HW34945, HW34954, HW34960, HW35171, HW35228, HW35230, HW35278, HW35282, HW35283, HW35297, HW35298, HW35300, HW35303, HW35304, HW35306, HW35307, HW35512, HW35513, HW35514, HW35516, HW35517, HW35518, HW35519, HW35520, HW35521, HW35522, HW35523, HW35525, HW35526, HW35527, HW35528, HW35565, HW35567, HW35568, HW35569, HW35570, HW35593, HW35600, HW35612, HW35614, HW35615, HW35616, HW35617, HW35618, HW35645, HW35650, HW35652, HW35685, HW35725, HW35734, HW35737, HW35738, HW35740, HW35742, HW35743, HW35744, HW35745, HW35746, HW35747, HW35748, HW35751, HW35752, HW35760, HW35762, HW35765 GTIN 00888707002462, Serial Numbers: HW31586, HW31700, HW32035 GTIN 00888707002424, Serial Numbers: HW31698, HW31851 GTIN 00888707002448, Serial Numbers: HW31779, HW31780, HW32546, HW34095, HW34617, HW35398, HW35511, HW35589, HW35597 GTIN 00888707002455, Serial Numbers: HW31852, HW31856, HW32025, HW32198, HW34026, HW34637, HW35544 GTIN 00888707006019, Serial Numbers: HW31983, HW34720, HW34974, HW35376, HW35377, HW35849, HW35866, HW35869, HW35871, HW35873, HW35877, HW35878, HW35879, HW35880, HW35998, HW36000, HW36056, HW36057, HW36058, HW36059, HW36060, HW36061, HW36062, HW36063, HW36113, HW36114, HW36115, HW36116, HW36117, HW36118, HW36119, HW36120, HW36145, HW36146, HW36155, HW36156, HW36157, HW36158, HW36159, HW36167, HW36221, HW36222, HW36223, HW36350, HW36351, HW36352 GTIN 00888707006408, Serial Numbers: HW34292, HW35932, HW36083, HW36084, HW36087, HW36129, HW36130, HW36171, HW36172, HW36173, HW36212, HW36247, HW36250, HW36261, HW36263, HW36264, HW36265, HW36266, HW36348, HW36349, HW36413, HW36414, HW36415, HW36416, HW36594, HW36638 GTIN 00888707002554, Serial Numbers: HW34337, HW34362, HW34484, HW34591, HW34594, HW34625, HW34682, HW34693, HW34732, HW34833, HW34838, HW34846, HW34876, HW34884, HW34889, HW35052, HW35057, HW35061, HW35067, HW35316, HW35323, HW35359, HW35362, HW35363, HW35364, HW35458, HW35536, HW35585, HW35620, HW35622, HW35623, HW35709, HW35710, HW35728, HW35732, HW35758, HW35759, HW35764, HW35767, HW35771, HW35772, HW35775, HW35777, HW35780, HW35782 GTIN 00888707004558, Serial Numbers: HW34759, HW34785, HW35655, HW35656 GTIN 00888707002141, Serial Numbers: HW34848, HW34849, HW35711 GTIN 00888707006323, Serial Numbers: HW34982, HW35031, HW35379, HW35380, HW35381, HW35382, HW35383, HW35384, HW35388, HW35389, HW35390, HW35391, HW35392, HW35621, HW35642, HW35654, HW35712, HW35722, HW35727, HW35729, HW35736, HW35755, HW35770, HW35781, HW35783, HW35977, HW36024, HW36025, HW36027, HW36028, HW36029, HW36030, HW36032, HW36033, HW36034, HW36042, HW36043, HW36044, HW36045, HW36046, HW36064, HW36065, HW36072, HW36073, HW36074, HW36075, HW36076, HW36077, HW36078, HW36079, HW36080, HW36081, HW36082, HW36123, HW36124, HW36125, HW36126, HW36127, HW36128, HW36133, HW36134, HW36135, HW36136, HW36137, HW36139, HW36140, HW36141, HW36142, HW36143, HW36144, HW36147, HW36148, HW36149, HW36150, HW36151, HW36152, HW36153, HW36187, HW36188, HW36189, HW36190, HW36191, HW36202, HW36203, HW36204, HW36206, HW36207, HW36208, HW36209, HW36210, HW36234, HW36235, HW36236, HW36237, HW36238, HW36239, HW36242, HW36244, HW36245, HW36267, HW36268, HW36271, HW36272, HW36274, HW36275, HW36284, HW36286, HW36290, HW36291, HW36292, HW36293, HW36295, HW36296, HW36297, HW36339, HW36340, HW36342, HW36343, HW36344, HW36346, HW36364, HW36366, HW36367, HW36375, HW36378, HW36379, HW36380, HW36439, HW36440, HW36441, HW36443, HW36444, HW36445, HW36446, HW36450, HW36451, HW36452, HW36453, HW36454, HW36455, HW36456, HW36520, HW36521, HW36522, HW36524, HW36528, HW36529, HW36532, HW36533, HW36534, HW36535, HW36537, HW36539, HW36548, HW36549, HW36550, HW36551, HW36552, HW36604, HW36605, HW36606, HW36607, HW36608, HW36639, HW36640, HW36641, HW36645, HW36649, HW36651, HW36661, HW36674, HW36675, HW36676, HW36687, HW36714, HW36729, HW36731, HW36732, HW36733, HW36734, HW36735, HW36737, HW36738, HW36739, HW36744, HW36908, HW36909, HW36922, HW36923, HW36925, HW36932 GTIN 00888707005951, Serial Numbers: HW35279, HW35331, HW35503, HW35524, HW35581, HW35675, HW35715, HW35720, HW35739, HW35768, HW35787, HW35788, HW35789, HW35790, HW35791, HW35792, HW35793, HW35794, HW35795, HW35796, HW35797, HW35798, HW35799, HW35821, HW35822, HW35823, HW35824, HW35825, HW35826, HW35827, HW35831, HW35833, HW35834, HW35835, HW35836, HW35837, HW35838, HW35839, HW35840, HW35841, HW35842, HW35843, HW35844, HW35845, HW35846, HW35883, HW35887, HW35888, HW35889, HW35890, HW35891, HW35892, HW35893, HW35894, HW35895, HW35896, HW35897, HW35898, HW35900, HW35905, HW35906, HW35907, HW35921, HW35925, HW35926, HW35928, HW35930, HW35933, HW35934, HW35945, HW35946, HW35947, HW35948, HW35949, HW35950, HW35951, HW35952, HW35953, HW35954, HW35955, HW35956, HW35957, HW35959, HW35960, HW35961 GTIN 00888707002264, Serial Numbers: HW35741, HW35773 GTIN 00888707006316, Serial Numbers: HW35754, HW35774, HW35865, HW35868, HW35874, HW35940, HW35958, HW35962, HW35963, HW35964, HW35965 GTIN 00888707005968, Serial Numbers: HW35801, HW35802, HW35803, HW35804, HW35805, HW35806, HW35807, HW35808, HW35809, HW35810, HW35811, HW35812, HW35813, HW35814, HW35815, HW35816, HW35817, HW35818, HW35819, HW35820, HW35847, HW35848, HW35850, HW35851, HW35852, HW35853, HW35854, HW35855, HW35856, HW35857, HW35858, HW35859, HW35860, HW35861, HW35862, HW35863, HW35864, HW35867, HW35870, HW35875, HW35923, HW35924, HW35927, HW35929, HW35931, HW35935, HW35936, HW35937, HW35938, HW35939, HW35966, HW35967, HW35968, HW35969, HW35970, HW35971, HW35972, HW35973, HW35975, HW35976, HW35978, HW35988, HW35989, HW35990, HW35991, HW35993, HW35994, HW35995, HW35996, HW35997 GTIN 00888707006309, Serial Numbers: HW35876, HW35882, HW35884, HW35885, HW35886, HW35941, HW35942, HW35999 GTIN 00888707006347, Serial Numbers: HW35902, HW35903, HW35908, HW35909, HW35910, HW35911, HW35912, HW35913, HW35914, HW35915, HW35918, HW35920, HW36001, HW36002, HW36003, HW36016, HW36035, HW36036, HW36037, HW36038, HW36039, HW36040, HW36048, HW36049, HW36050, HW36051, HW36052, HW36053, HW36054, HW36055, HW36088, HW36089, HW36092, HW36093, HW36094, HW36095, HW36096, HW36097, HW36100, HW36101, HW36102, HW36103, HW36105, HW36106, HW36107, HW36108, HW36109, HW36110, HW36111, HW36112, HW36170, HW36540, HW36541, HW36542, HW36543, HW36544, HW36565, HW36566, HW36568, HW36569, HW36570, HW36571, HW36572, HW36573, HW36574, HW36575, HW36576, HW36577, HW36578, HW36579, HW36581, HW36582, HW36583, HW36584, HW36585, HW36586, HW36611, HW36612, HW36615, HW36616, HW36618, HW36619, HW36620, HW36621, HW36622, HW36623 GTIN 00888707006446, Serial Numbers: HW35979, HW35980, HW35982, HW35984, HW36160, HW36161, HW36162, HW36163, HW36164, HW36278, HW36287, HW36381, HW36382, HW36383, HW36384, HW36385, HW36386, HW36387, HW36388, HW36460, HW36462, HW36464, HW36465, HW36469, HW36470, HW36471, HW36472, HW36480, HW36481, HW36482, HW36483, HW36484, HW36486, HW36487, HW36489, HW36491, HW36492, HW36494, HW36495 GTIN 00888707006392, Serial Numbers: HW35986, HW35987, HW36066, HW36085, HW36086, HW36213, HW36214, HW36215, HW36216, HW36217, HW36218, HW36219, HW36220, HW36289, HW36311, HW36417, HW36418, HW36517, HW36587, HW36588, HW36589, HW36590, HW36652, HW36656, HW36860 GTIN 00888707007290, Serial Numbers: HW36421, HW36422 GTIN 00888707006354, Serial Numbers: HW36838, HW36880 No GTIN, Serial Numbers: HW26099, HW26126, HW26330, HW26660, HW26939, HW26940, HW27099, HW27452, HW27678, HW28063, HW28199, HW28373, HW28418, HW28895, HW28973, HW29073, HW29214, HW29298, HW29318, HW29351, HW29353, HW29354, HW29359

Distribution

Worldwide Distribution

Quantity

2225 Units