FDA Enforcement
Class II
Terminated
Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).
Recall: Z-2055-2017
·
Reported May 17, 2017
Enforcement
- Recall Number
- Z-2055-2017
- Event ID
- 77120
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Greatbatch Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 17, 2017
- Initiation Date
- April 23, 2017
- Classification Date
- May 9, 2017
- Termination Date
- March 7, 2019
- Address
- 2300 Berkshire Ln N, N/A, Plymouth, MN, 55441-4575, United States
Description
Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).
Reason
Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.
Code Info
Lot numbers : Linear Straight Broach Handle (Rasp Handle): 2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 3433392, 3332697. EZ Clean" Linear Broach Handle: 3508491.
Distribution
US Distribution to the state of : TX
Quantity
205