FDA Enforcement Class II Ongoing

Canon Vantage Titan 3T, Premium Open-Bore MRI System, REF: MRT-3010 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

Recall: Z-2046-2021 · Reported July 21, 2021

Enforcement

Recall Number
Z-2046-2021
Event ID
88014
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Canon Medical System, USA, INC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 21, 2021
Initiation Date
May 21, 2021
Classification Date
July 9, 2021
Address
2441 Michelle Dr, Tustin, CA, 92780-7047, United States

Description

Canon Vantage Titan 3T, Premium Open-Bore MRI System, REF: MRT-3010 - Product Usage: use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

Reason

The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

Code Info

Affected Serial Numbers: A5B1182002, A5B11Y2005, A5D14X2032, A5A1082001, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5F1562034, A5B1192004, A5C1222010, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5C1232011, A5C1252012, A5D13Z2027, A5D1492030, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5G1652037, A5D1362021, A5C1292015, A5D13X2024, A5C1422029, A5D1322020, A5B11Z2007, A5C1212008, A5C12Z2018, A5D1372022

Distribution

US Nationwide distributions in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY.