FDA Enforcement Class II Terminated

nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicTumorEx software enables processing and visualization of dynamic MR imaging datasets of the brain, showing properties of changes in contrast overtime. It generates various parametric images calculated from the image intensity variations and provides tools to extract and visualize such parametric properties within specified volumes of interest. Structural datasets providing anatomical information can be sued to evaluate the extent of a specified sub-volume, such as a tumor. Comparison of imaging studies performed at different study dates may also be performed to support the diagnostic process.

Recall: Z-2042-2018 · Reported June 6, 2018

Enforcement

Recall Number
Z-2042-2018
Event ID
78460
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
NordicNeuroLab AS
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 6, 2018
Initiation Date
December 15, 2014
Classification Date
May 31, 2018
Termination Date
March 24, 2021
Address
Mollendalsveien 65c, N/A, Bergen, N/A, N/A, Norway

Description

nordicTumorEx 1.0 nordicTumorEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicTumorEx software enables processing and visualization of dynamic MR imaging datasets of the brain, showing properties of changes in contrast overtime. It generates various parametric images calculated from the image intensity variations and provides tools to extract and visualize such parametric properties within specified volumes of interest. Structural datasets providing anatomical information can be sued to evaluate the extent of a specified sub-volume, such as a tumor. Comparison of imaging studies performed at different study dates may also be performed to support the diagnostic process.

Reason

An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.

Code Info

Version 1.0.4

Distribution

Worldwide Distribution - USA (nationwide) and to the countries of : SA, Russia, South Korea, UK, Japan, Australia, Sweden, Norway, Greece

Quantity

16 licenses