FDA Enforcement Class II Terminated

LINX Reflux Management System, Implant Clasp, 13-16 Bead, 1.5T, Product Code LXM-xx, sterile, Rx only. Labeled as: a. Product Code LXM-13; b. Product Code LXM-14; c. Product Code LXM-15; d. Product Code LXM-16; Laparoscopic, fundic-sparing anti-reflux procedure indicated for patients diagnosed with Gastroesophageal Reflux Disease (GERD) as defined by abnormal pH testing, and who are seeking an alternative to continuous acid suppression therapy (i.e. proton pump inhibitors or equivalent) in the management of their GERD.

Recall: Z-2041-2018 · Reported June 6, 2018

Enforcement

Recall Number
Z-2041-2018
Event ID
79822
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Torax Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 6, 2018
Initiation Date
April 20, 2018
Classification Date
May 31, 2018
Termination Date
November 4, 2020
Address
4188 Lexington Ave N, N/A, Shoreview, MN, 55126-6106, United States

Description

LINX Reflux Management System, Implant Clasp, 13-16 Bead, 1.5T, Product Code LXM-xx, sterile, Rx only. Labeled as: a. Product Code LXM-13; b. Product Code LXM-14; c. Product Code LXM-15; d. Product Code LXM-16; Laparoscopic, fundic-sparing anti-reflux procedure indicated for patients diagnosed with Gastroesophageal Reflux Disease (GERD) as defined by abnormal pH testing, and who are seeking an alternative to continuous acid suppression therapy (i.e. proton pump inhibitors or equivalent) in the management of their GERD.

Reason

Possible out of specification condition which may allow a bead component to separate from an adjacent wire link. This condition may result in a discontinuous or open LINX device.

Code Info

a. Product Code LXM-13, Part No. 4549-13, GTIN 00855106005394; b. Product Code LXM-14, Part No. 4549-14, GTIN 00855106005400; c. Product Code LXM-15, Part No. 4549-15, GTIN 00855106005417; d. Product Code LXM-16, Part No. 4549-16, GTIN 00855106005424; Lot No. 6100 through 14055, 15286, 15287, 15314, 15315, 15317, and 16557

Distribution

Worldwide Distribution: US (nationwide) in states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY; and in countries of: Austria, France, Germany, Italy, Scotland, Spain, Switzerland, United Kingdom.

Quantity

9,131 devices in total