FDA Enforcement Class II Ongoing

Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915

Recall: Z-2038-2025 · Reported July 9, 2025

Enforcement

Recall Number
Z-2038-2025
Event ID
96946
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
July 9, 2025
Initiation Date
May 9, 2025
Classification Date
June 27, 2025
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101, United States

Description

Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915

Reason

A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.

Code Info

VITROS 5600 Integrated System; Product Code: 6802413; UDI: 10758750002740; VITROS 5600 Integrated System - Certified/Refurbished; Product Code: 6802915; UDI: 10758750002740; Range of installed serial numbers: 56000143-56005096 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802915 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process.

Distribution

Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

Quantity

3117 units (813 US, 2304 OUS)