Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU DYNJ64138B; 3. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572A; 4. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572B; 5. MICROSURGERY PACK, Medline Kit Number/SKU DYNJ67284A; 6. BASIC EYE PACK, Medline Kit Number/SKU DYNJ81411A; 7. SMR SINUS PACK, Medline Kit Number/SKU DYNJ84478; 8. FESS/SEPTO PACK, Medline Kit Number/SKU DYNJ85025; 9. PLASTIC EYE PACK, Medline Kit Number/SKU DYNJ89030; 10. NEURO, Medline Kit Number/SKU DYNJ902935K; 11. OCULAR, Medline Kit Number/SKU DYNJ909065; 12. OCULAR, Medline Kit Number/SKU DYNJ909065A.
Enforcement
- Recall Number
- Z-2037-2026
- Event ID
- 98664
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 6, 2026
- Initiation Date
- March 19, 2026
- Classification Date
- April 30, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU DYNJ64138B; 3. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572A; 4. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572B; 5. MICROSURGERY PACK, Medline Kit Number/SKU DYNJ67284A; 6. BASIC EYE PACK, Medline Kit Number/SKU DYNJ81411A; 7. SMR SINUS PACK, Medline Kit Number/SKU DYNJ84478; 8. FESS/SEPTO PACK, Medline Kit Number/SKU DYNJ85025; 9. PLASTIC EYE PACK, Medline Kit Number/SKU DYNJ89030; 10. NEURO, Medline Kit Number/SKU DYNJ902935K; 11. OCULAR, Medline Kit Number/SKU DYNJ909065; 12. OCULAR, Medline Kit Number/SKU DYNJ909065A.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Medline Kit Number/SKU DYNJ48725A: UDI/DI each 10193489853896, UDI/DI case 40193489853897, Lot Number: 24IMG757; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22FME241; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22DMC340; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21JMC829; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21IMC497; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21HMD139; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21GME747; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21FMA440; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21DBU602; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21CBJ936; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23IBC227; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23FBA458; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23DBC239; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23ABI841; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBP260; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBO842; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBN801; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22GBL794; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 24HBB998; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 24DBQ664; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 24CBQ260; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 26ABC645; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25LBB014; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25JBM975; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25HBN493; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25GBM010; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25FBJ206; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25CBN002; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25ABF979; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25ABA086; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 24JBE583; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 24JBG667; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 25KBQ749; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 25JBF743; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 25IBA573; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24IBO985; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24IBB916; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24GBY148; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24EBO714; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 23LBH320; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23KBO017; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23JBA275; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23GBK387; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23CBD302; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 26ABF562; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 25GBF664; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 25EBB838; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 25BBQ021; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24LBB454; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24JBE139; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24GBI010; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24FBR965; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24DBQ797; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24BBH448; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23LBL498; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23KBX879; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 23IBE948; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 23FBB413; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 24ABN367; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 24ABA089; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 23KBU302; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 25LBL874; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 25HBD880; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 25EBI919; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 24LBG184; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 24LBF578; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 24JBS764; Medline Kit Number/SKU DYNJ89030: UDI/DI each 10198459229725, UDI/DI case 40198459229726, Lot Number: 25JBH439; Medline Kit Number/SKU DYNJ89030: UDI/DI each 10198459229725, UDI/DI case 40198459229726, Lot Number: 25ABP922; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23IBW005; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23IBG214; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23FBP153; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23FBN621; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23EBW204; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23EBE141; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 24DBP594; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 24BBB058; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 24BBB100; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23LBE522; Medline Kit Number/SKU DYNJ909065: UDI/DI each 10195327260019, UDI/DI case 40195327260010, Lot Number: 23CMB048; Medline Kit Number/SKU DYNJ909065: UDI/DI each 10195327260019, UDI/DI case 40195327260010, Lot Number: 23CMA495; Medline Kit Number/SKU DYNJ909065: UDI/DI each 10195327260019, UDI/DI case 40195327260010, Lot Number: 22LMD749; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 23HMI877; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 23HMH413; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 25GMJ496; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 25DMD357; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 25AMJ130; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24IME711; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24HME507; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24DMF364; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24CMI687; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24CMG984; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24BMI425.
Worldwide distribution.
3290 kits