Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Recall: Z-2037-2012 · Reported July 25, 2012

Enforcement

Recall Number
Z-2037-2012
Event ID
62372
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hitachi Medical Systems America Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 25, 2012
Initiation Date
May 4, 2012
Classification Date
July 19, 2012
Termination Date
February 19, 2013
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087, United States

Description

Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Reason

Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.

Code Info

Serial #'s S5001 - S5011

Distribution

US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.

Quantity

Six (6) units