Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Enforcement
- Recall Number
- Z-2037-2012
- Event ID
- 62372
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hitachi Medical Systems America Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 25, 2012
- Initiation Date
- May 4, 2012
- Classification Date
- July 19, 2012
- Termination Date
- February 19, 2013
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087, United States
Description
Hitachi Scenaria CT System Software Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.
Serial #'s S5001 - S5011
US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.
Six (6) units