Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.
Enforcement
- Recall Number
- Z-2036-2014
- Event ID
- 68696
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Steris Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Visit
- Report Date
- July 16, 2014
- Initiation Date
- June 30, 2014
- Classification Date
- July 10, 2014
- Termination Date
- July 1, 2015
- Address
- 5960 Heisley Rd, Mentor, OH, 44060-1834, United States
Description
Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.
STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.
Serial Number Range: 3632511020 - 3613214010
US Nationwide Distribution - in the states of AL, AR, CA, DE, FL, GA, IL, IA, LA, MA, MI, NJ, NY, OH, OK, OR, SC, TX, VA. and in the country of Canada.