FDA Enforcement Class II Terminated

Artis zeego, Model Number 10280959

Recall: Z-2034-2019 · Reported July 31, 2019

Enforcement

Recall Number
Z-2034-2019
Event ID
83203
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2019
Initiation Date
June 12, 2019
Classification Date
July 22, 2019
Termination Date
September 18, 2020
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

Artis zeego, Model Number 10280959

Reason

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Code Info

Serial Numbers: 160395 160484 160502 160351 160392 160449 160463 160313 160411 160406 160106 160134 160136 160347 160312 160488 160448 160427 160455 160486 160435 160451 160346 160384 160100 160493 160391 160321 160376 160447 160471 160481 160401 160119 160403 160404 160501 160438 160388 160378 160419 160114 160450 160429 160149 160137 160152 160153 160154 160138 160446 160368 160394 160412 160387 160432 160147 160433 160483 160335 160402 160356 160425 160148 160337 160460 160485 160443 160461 160370 160381 160508 160122 160407 160325 160494 160390 160507 160144 160487 160418 160007 160001 160363 160341 160139 160430 160482 160006 160462 160520 160505 160445 160322 160323 160382 160421 160474 160116 160428 160131 160318 160357 160369 160498 160413 160414 160128 160373 160509 160551 160327 160338 160477 160519 160141 160469

Distribution

US Nationwide Distribution

Quantity

117