Stryker AVS Navigator Trials 6mm-15mm Manufactured by: Stryker Spine SAS Z.1 Marticot-33610-Cestas -France +33 (0) 5.57.97.06.30 http;//www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Court. Allendale NJ 07401-1677 USA +1-201-780-8000 Cestas France 33610. Intended Use: The trials are designed and intended to measure the height of the disc space intra-operatively.
Enforcement
- Recall Number
- Z-2033-2012
- Event ID
- 61723
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Spine
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 25, 2012
- Initiation Date
- January 13, 2012
- Classification Date
- July 19, 2012
- Termination Date
- July 29, 2013
- Address
- 2 Pearl Ct, Allendale, NJ, 07401-1611, United States
Description
Stryker AVS Navigator Trials 6mm-15mm Manufactured by: Stryker Spine SAS Z.1 Marticot-33610-Cestas -France +33 (0) 5.57.97.06.30 http;//www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Court. Allendale NJ 07401-1677 USA +1-201-780-8000 Cestas France 33610. Intended Use: The trials are designed and intended to measure the height of the disc space intra-operatively.
Beginning in February 2011, Stryker Spine began receiving reports related to the shaft of the AVS Navigator trial failing at the junction of the trial head.
Class I; 510(k) Exempt Description: AVS¿ Navigator Trials. Size 6mm -15mm. Catalog Number: 48392206,48392207,48392208,48392209,48392210, 48392211,48392212,48392213,48392214,48392215. Lot Code: 104864,104866, 104867, 104868, 104869, 104870, 10E331, 104871, 104872, 104873, 104874, 104875.
Nationwide.
273 units