FDA Enforcement Class I Ongoing

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Recall: Z-2032-2025 · Reported July 16, 2025

Enforcement

Recall Number
Z-2032-2025
Event ID
96856
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 16, 2025
Initiation Date
May 30, 2025
Classification Date
July 9, 2025
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Reason

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Code Info

All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025

Distribution

Worldwide

Quantity