FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096
Enforcement
- Recall Number
- Z-2030-2021
- Event ID
- 88064
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- The Binding Site Group, Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 14, 2021
- Initiation Date
- May 20, 2021
- Classification Date
- July 7, 2021
- Termination Date
- February 4, 2025
- Address
- 8 Calthorpe Road, Birmingham, N/A, United Kingdom
Description
FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096
The high and low level "Quality Controls" demonstrate a positive bias of approximately 10% compared to assigned values.
Lot Numbers: 487719/ Exp: 31-Jul-2022, 480133/ Exp: 30-Sep-2022, Lot 457890/ Exp: 30-Sep-2022
Worldwide distribution - US Nationwide distribution in the states of CA, GA, FL, MA, MO, MI, WA, NY, PA, MN, AL, MD, OK, NE, IN, ID, OH, TX, HI, DC and the countries of Canada, Australia, Austria, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, Denmark, Estonia, France, Germany, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Spain, Switzerland, United Kingdom, Argentina, Egypt, GUADELOUPE, HONG KONG, India, Iran, Israel, Korea, Lebanon, Malaysia, Morocco, Oman, Singapore, Taiwan, Tunisia, Vietnam.