FDA Enforcement Class II Ongoing

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060J; 5. NEURO BASIN KIT, Medline Kit Number/SKU DYNJ80200B; 6. NEURO-SPINE BASIN PACK, Medline Kit Number/SKU DYNJ83019; 7. ENT PACK, Medline Kit Number/SKU DYNJ86328; 8. EDMOND SPINE PACK, Medline Kit Number/SKU DYNJ89263; 9. ENT PACK, Medline Kit Number/SKU DYNJ89675; 10. ENT PACK, Medline Kit Number/SKU DYNJ89675A; 11. ENT/PAIN PACK, Medline Kit Number/SKU DYNJT7493.

Recall: Z-2026-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-2026-2026
Event ID
98664
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 6, 2026
Initiation Date
March 19, 2026
Classification Date
April 30, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060J; 5. NEURO BASIN KIT, Medline Kit Number/SKU DYNJ80200B; 6. NEURO-SPINE BASIN PACK, Medline Kit Number/SKU DYNJ83019; 7. ENT PACK, Medline Kit Number/SKU DYNJ86328; 8. EDMOND SPINE PACK, Medline Kit Number/SKU DYNJ89263; 9. ENT PACK, Medline Kit Number/SKU DYNJ89675; 10. ENT PACK, Medline Kit Number/SKU DYNJ89675A; 11. ENT/PAIN PACK, Medline Kit Number/SKU DYNJT7493.

Reason

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Code Info

Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25IMD003; Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25HMH643; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22GBS127; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22FBB472; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25FBE752; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25EBA260; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25DBG176; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25CBD819; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25BBA593; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 24KBJ163; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23KMI415; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMK023; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMB211; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23IMC591; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23HMG291; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23GMG136; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23FME285; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23EME227; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23CMD903; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMI915; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMH419; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24JMG571; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24HMF066; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24FMG130; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24CMI441; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24BMD229; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24AMH016; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26BMB944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26AMJ197; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25KMC587; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25JMD275; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25IMC944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25GMF811; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25FME706; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25EMA516; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25CME296; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 24LMD726; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 22HBH553; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 26BBF722; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25JBB835; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBO944; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBL998; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBH356; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBF751; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25DBB435; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25BBC547; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24KBR868; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24JBV969; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBB952; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBC125; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 22JBR211; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25LBS635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25KBE004; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25HBR489; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25FBL484; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25DBD392; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25ABD635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24KBB462; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24JBE518; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24GBE890; Medline Kit Number/SKU DYNJ89263: UDI/DI each 10198459253348, UDI/DI case 40198459253349, Lot Number: 25BBK629; Medline Kit Number/SKU DYNJ89675: UDI/DI each 10198459281693, UDI/DI case 40198459281694, Lot Number: 25CME067; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMG054; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMA771; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 25GME903; Medline Kit Number/SKU DYNJT7493: UDI/DI each 10198459626258, UDI/DI case 40198459626259, Lot Number: 26ABA254.

Distribution

Worldwide distribution.

Quantity

9805 kits