Medline medical procedure kits, containing Medline Neuro Sponges (102 in total)
Enforcement
- Recall Number
- Z-2025-2026
- Event ID
- 98664
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 6, 2026
- Initiation Date
- March 19, 2026
- Classification Date
- April 30, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline medical procedure kits, containing Medline Neuro Sponges (102 in total)
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23KBN711; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23GBD593; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23GBD781; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23DBF906; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22LBI985; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22KBC504; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22IBF741; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22FBX069; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22DBS948; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22CBX667; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 21IBT137; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 21HBG608; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 25GBV215; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 25GBA852; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24KBJ988; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24JBC528; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24IBV131; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24IBP386; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24GBA976; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24CBU492; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23KBW165; Medline Kit Number/SKU CDS980375M: UDI/DI each 10198459447327, UDI/DI case 40198459447328, Lot Number: 26ABK337; Medline Kit Number/SKU CDS980375M: UDI/DI each 10198459447327, UDI/DI case 40198459447328, Lot Number: 25JBE079; Medline Kit Number/SKU CDS980375M: UDI/DI each 10198459447327, UDI/DI case 40198459447328, Lot Number: 25HBV443; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25IBR604; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25HBR025; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25GBB389; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25FBC324; Medline Kit Number/SKU CDS983100D: UDI/DI each 10198459007934, UDI/DI case 40198459007935, Lot Number: 24KBH657; Medline Kit Number/SKU CDS983100D: UDI/DI each 10198459007934, UDI/DI case 40198459007935, Lot Number: 24GBN523; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23IBT936; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23HBA798; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23FBF767; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23DBD282; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 22KBJ790; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 22HBD313; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 24DBC806; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 24BBN518; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23LBL541; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23LBD588; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23HDB332; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23GDA495; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23FDB617; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23EDB945; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 22JDA727; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 22FDB173; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21JDB599; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21JDB403; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21IDC321; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21IDB329; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21HDA478; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21GDC463; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21FDA896; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21EDB291; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21CDC134; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21CDB258; Medline Kit Number/SKU CDS983493M: UDI/DI each 10195327497415, UDI/DI case 40195327497416, Lot Number: 24ADA328; Medline Kit Number/SKU CDS983493M: UDI/DI each 10195327497415, UDI/DI case 40195327497416, Lot Number: 23LDA144; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 26CDA415; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 26BDA065; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 25LDA842; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 25IDB325; Medline Kit Number/SKU DYNJ0395085P: UDI/DI each 10193489432718, UDI/DI case 40193489432719, Lot Number: 21DKA483; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23IMD698; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23GMB527; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23FMF102; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23EMH628; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23CMA614; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23BME548; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22JMB623; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22GMG727; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22FME591; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22DMG455; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 26AMA936; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25KMG652; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25JME572; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25GMF834; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25EMA958; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25DME904; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25CME519; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24LMF057; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24JMH001; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24IMC603; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24HMA307; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24EMH324; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24EMA549; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24AMJ677; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23LMC624; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23IMG288; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23GMF398; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23FMH043; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23FMD596; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23EMF866; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23DMH874; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23CMH139; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23CME193; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23BMD471; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22LMD522; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22HMG591; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22GME991; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22GME992; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22DMC069; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22OBD206; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22CBF109; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 21IBM538; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 21IBM610; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 21FBP997; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24JMH447; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24IMF995; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24FMI942; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24EME920; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24DMJ136; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23LMH895; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23KMC601; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23GMC715; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23DME468; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23BMC887; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 21LLA690; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 24BMA691; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 24AMH656; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23LMB786; Medline Kit Number/SKU DYNJ39211F: UDI/DI each 10195327313654, UDI/DI case 40195327313655, Lot Number: 23BMD216; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 22AMG375; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21KMG254; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21HMB060; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21FMD643; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21DME180; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23IBB468; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23FBO732; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23EBV753; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23EBQ610; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23DBM685; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23ABH385; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 22GBK343; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 22EBB123; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 22ABE215; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 21ABM833; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24JBE450; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24JBF281; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24JBA800; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24IBT612; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24FBS381; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23LBN005; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 26BBK430; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 25JBU219; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 25GBW762; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 25DBU105; Medline Kit Number/SKU DYNJ43859J: UDI/DI each 10193489893700, UDI/DI case 40193489893701, Lot Number: 23EME342; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22IBI991; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22EBP134; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22ABX616; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22ABA063; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 21JBD678; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 21HBQ587; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 21DBC693; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 23GME117; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 23EMI284; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 23AMJ001; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22KMD020; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22GBW975; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22FBW091; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22OBJ071; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22CBC147; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 21VBD092; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 23HMG668; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24JMC633; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24IMI654; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24IMB838; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24HMB162; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24FMG045; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24EMD442; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24DME855; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24CMH418; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24BMK020; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24AMG855; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24AMB928; Medline Kit Number/SKU DYNJ46504F: UDI/DI each 10198459140761, UDI/DI case 40198459140762, Lot Number: 25HMI601; Medline Kit Number/SKU DYNJ46504F: UDI/DI each 10198459140761, UDI/DI case 40198459140762, Lot Number: 25GMG101; Medline Kit Number/SKU DYNJ46504F: UDI/DI each 10198459140761, UDI/DI case 40198459140762, Lot Number: 25EMF342; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 22JBA979; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 26BBR125; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 25KBQ493; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24JBD583; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24HBA712; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24EBM092; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24DBK313; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24BBO353; Medline Kit Number/SKU DYNJ48191G: UDI/DI each 10193489476606, UDI/DI case 40193489476607, Lot Number: 21JMH268; Medline Kit Number/SKU DYNJ48191G: UDI/DI each 10193489476606, UDI/DI case 40193489476607, Lot Number: 21JMF238; Medline Kit Number/SKU DYNJ48191G: UDI/DI each 10193489476606, UDI/DI case 40193489476607, Lot Number: 21DMB033; Medline Kit Number/SKU DYNJ48191I: UDI/DI each 10195327057817, UDI/DI case 40195327057818, Lot Number: 22EME800; Medline Kit Number/SKU DYNJ48191I: UDI/DI each 10195327057817, UDI/DI case 40195327057818, Lot Number: 22CMD031; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 23EMA429; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 23BMD353; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 22LMH655; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 22JME786; Medline Kit Number/SKU DYNJ48191K: UDI/DI each 10195327437541, UDI/DI case 40195327437542, Lot Number: 23GMG423; Medline Kit Number/SKU DYNJ48191K: UDI/DI each 10195327437541, UDI/DI case 40195327437542, Lot Number: 23LMH905; Medline Kit Number/SKU DYNJ48191L: UDI/DI each 10195327630690, UDI/DI case 40195327630691, Lot Number: 24EMH956; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25GMD380; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25EMA899; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25BMD703; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25BMB374; Medline Kit Number/SKU DYNJ48191O: UDI/DI each 10198459602641, UDI/DI case 40198459602642, Lot Number: 26AMB007; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21JMD339; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21FMF721; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21FME498; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21EMA403; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21HBG397; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21FBE715; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21CBW195; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21ABG033; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21ABH228; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23JBY274; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23FBM738; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23EBS795; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23DBQ176; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23DBB931; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23CBB075; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23BBC518; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22JBD638; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22FBJ435; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22EBL603; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22CBV860; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22CBP225; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 21LBU279; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 21KBP799; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 26CBB708; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 26ABM764; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25LBQ922; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25KBI476; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25IBS238; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25IBB148; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25HBM494; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25FBI994; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25DBU225; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25BBE810; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25ABK916; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24LBB828; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24JBX741; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24IBR297; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24HBI405; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24FBP312; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24EBN616; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24CBL687; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24CBJ095; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24BBS591; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23IBD079; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23GBL994; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23EBS598; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23EBL984; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23DBH411; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23CBV513; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23BBD492; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 22DBQ354; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 22BBB405; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21KBW367; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21KBC969; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21FBE526; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21FBE506; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21DBM118; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21DBD151; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21BBA803; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21ABF228; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 26BBS083; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 25IBE859; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 25BBJ593; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 25BBJ601; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 24HBD973; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 24FBB019; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 24CBK963; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23KBX604; Medline Kit Number/SKU DYNJ59400K: UDI/DI each 10195327492489, UDI/DI case 40195327492480, Lot Number: 24GMC759; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 23IBC981; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 23CBB054; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 23ABK120; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 22GBU138; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 22BBT750; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 21JBZ859; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 21FBA944; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 21ABM493; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 26BBG778; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 4
0193489362061, Lot Number: 25KBN868; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 25EBF013; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24LBK254; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24HBM708; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24DBB730; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24ABC264; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23EMG712; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23EMA976; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23CMH213; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23BMH981; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23AMI112; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22JMF099; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22EMA459; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22DMB172; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22BMG965; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21LMG903; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21KMH623; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21JMH398; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21JMF247; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21HMD136; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21GMD160; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21FMC247; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23JMG294; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23IMC541; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23GMF129; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23FMI543; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 25CMG672; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 25AMB065; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24LMA593; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24JMI177; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24HMH536; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24HMD457; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24GMJ778; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24FMC742; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24EME114; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24CMI272; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23KMJ516; Medline Kit Number/SKU DYNJ60392F: UDI/DI each 10198459313424, UDI/DI case 40198459313425, Lot Number: 25EMH894; Medline Kit Number/SKU DYNJ60392F: UDI/DI each 10198459313424, UDI/DI case 40198459313425, Lot Number: 25DMJ856; Medline Kit Number/SKU DYNJ62636: UDI/DI each 10193489675108, UDI/DI case 40193489675109, Lot Number: 21AMA163; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 23KBT422; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 23GBF930; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 22HBQ384; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 22GBS133; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 26ABN799; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25KBN615; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25FBQ548; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25FBO124; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25FBF480; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 24LBP070; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 23LBN303; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22IBS344; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22GBC932; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22DBL674; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22DBB480; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22CBV253; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 21LBO403; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 22EBR214; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 22BBG688; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 22ABW065; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21KBK820; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21JBL906; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21FBR921; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21FBR939; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21EBA643; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21BBQ790; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21BBC028; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23JBE733; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23GBK608; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23DBJ226; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 22LBJ519; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 25GBL417; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 25EBC969; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24KBQ659; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24HBH157; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24FBM160; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24BBA851; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23LBL662; Medline Kit Number/SKU DYNJ64658B: UDI/DI each 10198459560675, UDI/DI case 40198459560676, Lot Number: 26CBA976; Medline Kit Number/SKU DYNJ64658B: UDI/DI each 10198459560675, UDI/DI case 40198459560676, Lot Number: 25LBF574; Medline Kit Number/SKU DYNJ65028: UDI/DI each 10193489240894, UDI/DI case 40193489240895, Lot Number: 21DBG884; Medline Kit Number/SKU DYNJ65028: UDI/DI each 10193489240894, UDI/DI case 40193489240895, Lot Number: 21BBQ227; Medline Kit Number/SKU DYNJ65028: UDI/DI each 10193489240894, UDI/DI case 40193489240895, Lot Number: 21ABC103; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22DMA079; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22CBA570; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22BBS875; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22BBH976; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21LBB949; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21JBG683; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21GBO340; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21GBL438; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21GLA583; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21ELA300; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21DLA305; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21DLA054; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21CLA170; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21CLA152; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 23DMA746; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 23BMC966; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22IMF531; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22GMG777; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22FMA171; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22CLA200; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22ALA499; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 21JLB002; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 21HLA839; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 23IMG461; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 23EMF925; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24JMA680; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24HMI731; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24HMG728; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24FMA661; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24BMC283; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24AMG558; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22GMA633; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22EMG944; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22EMA059; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22DMA733; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21KMF575; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21JMD129; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21HMB364; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21GMB819; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21EMD513; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21CMB774; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23JMJ037; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23IMC044; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23DMI656; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23BMH340; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24IME987; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24HMA605; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24GMA632; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24DMG133; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24CMK124; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24BMI337; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23LMD063; Medline Kit Number/SKU DYNJ81727: UDI/DI each 10195327137113, UDI/DI case 40195327137114, Lot Number: 22GMI084; Medline Kit Number/SKU DYNJ81727: UDI/DI each 10195327137113, UDI/DI case 40195327137114, Lot Number: 22FMH411; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23KME276; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23IMB927; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23BMG594; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23BMG926; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23BMG593; Medline Kit Number/SKU DYNJ81727F: UDI/DI each 10198459260940, UDI/DI case 40198459260941, Lot Number: 25CMA943; Medline Kit Number/SKU DYNJ81727F: UDI/DI each 10198459260940, UDI/DI case 40198459260941, Lot Number: 25BMH565; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23JDA488; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23FDC241; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23FDA551; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23DDB518; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 22GDB463; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23LDA469; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 26BME027; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25LMI396; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25LMH594; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25IMB428; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25EMG601; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23KMH124; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23FMJ566; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23FMF306; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23EME060; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23CMA243; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23AMC472; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 22IMH894; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 22IMG830; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 26BMA823; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25JMF655; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25HMC540; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25EMG428; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25BMG415; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24LME660; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24KMB725; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24HMJ214; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24EMA834; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24BMI629; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23LMD692; Medline Kit Number/SKU DYNJ83417: UDI/DI each 10195327265038, UDI/DI case 40195327265039, Lot Number: 22LBO285; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 23FBJ724; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 23EBI932; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 23EBI500; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 25DBJ446; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 25CBM860; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 25BBC119; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24JBF417; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24FBQ165; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24EBS240; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24DBK432; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24BBR722; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 26BBI958; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25JBO782; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25IBA640; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25GBD814; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25DBG181; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25CBG994; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 24LBS022; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 24CBQ298; Medline Kit Number/SKU DYNJ87503: UDI/DI each 10198459041822, UDI/DI case 40198459041823, Lot Number: 24LBJ556; Medline Kit Number/SKU DYNJ87503A: UDI/DI each 10198459337680, UDI/DI case 40198459337681, Lot Number: 25FBT430; Medline Kit Number/SKU DYNJ87503B: UDI/DI each 10198459563836, UDI/DI case 40198459563837, Lot Number: 25KBA102; Medline Kit Number/SKU DYNJ88201: UDI/DI each 10198459121524, UDI/DI case 40198459121525, Lot Number: 24JBF518; Medline Kit Number/SKU DYNJ88201A: UDI/DI each 10198459356360, UDI/DI case 40198459356361, Lot Number: 25IBJ505; Medline Kit Number/SKU DYNJ88201A: UDI/DI each 10198459356360, UDI/DI case 40198459356361, Lot Number: 25HBK860; Medline Kit Number/SKU DYNJ88201A: UDI/DI each 10198459356360, UDI/DI case 40198459356361, Lot Number: 25FBV160; Medline Kit Number/SKU DYNJ89375A: UDI/DI each 10198459720147, UDI/DI case 40198459720148, Lot Number: 26CMC035; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23IBH113; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23GBH719; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23GBH824; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23FBL900; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23EBW826; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23DBT401; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23CBQ543; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23CBE800; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23CBE801; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22LBS875; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22LBS876; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22HBV144; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22HBV582; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22GBZ303; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22GBS213; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22DBM741; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22DBC328; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21LBW483; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21KBK354; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21JBX680; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21IBR084; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21HBI219; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21HBB375; Medline Kit Number/SKU DYNJ901296A: UDI/DI each 10889942009992, UDI/DI case 40889942009993, Lot Number: 21DBE128; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 23FDB702; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 23FDA114; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 22GDA782; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 22FDB058; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21HDB858; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21HDA339; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21GDB427; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21GDB224; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21FDB291; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21DDB137; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21CDB779; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21BDA755; Medline Kit Number/SKU DYNJ901753K: UDI/DI each 10195327431211, UDI/DI case 40195327431212, Lot Number: 23IDA281; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 26ADB691; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25GDB518; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25GDA588; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25EDB366; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25CDB331; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 24IDB178; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 24GDA863; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 24BDA185; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 23LDB476; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23JBL937; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23FBV736; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23CBR362; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23ABC423; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22LBF244; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22HBS342; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22FBX915; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22EBR566; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22CBM758; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22ABF168; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21LBM300; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21KBZ611; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21IBJ211; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21IBJ212; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21EBH572; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21EBC781; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24KBA482; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24IBN252; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24GBN507; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24FBG214; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24FBE668; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24BBF142; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24ABC727; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23KBX358; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25LMH954; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25LMC207; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25HMG134; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25EMH523; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25DMA957; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25CMH640; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25BMD941; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25AMG740; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25AMA942; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 24LMC564; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 24IMG084; Medline Kit Number/SKU DYNJ904565D: UDI/DI each 10193489379150, UDI/DI case 40193489379151, Lot Number: 21FBD283; Medline Kit Nu mber/SKU DYNJ904565D: UDI/DI each 10193489379150, UDI/DI case 40193489379151, Lot Number: 21DBG641; Medline Kit Number/SKU DYNJ904565D: UDI/DI each 10193489379150, UDI/DI case 40193489379151, Lot Number: 21CBK121; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23DBF757; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23CBG746; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23ABS372; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23ABI406; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22IBO075; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22IBN313; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22IBN315; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22GBE006; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22GBE793; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22EBQ475; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22DBG440; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22CBP145; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 21LBE727; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 21HBM868; Medline Kit Number/SKU DYNJ904565G: UDI/DI each 10195327371784, UDI/DI case 40195327371785, Lot Number: 23DBP675; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23JBY083; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23JBC192; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23IBH848; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25FBM817; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25CBH988; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25ABW505; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25ABC626; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24KBG651; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24IBP291; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24HBP497; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24GBR801; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24FBS752; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24FBC560; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24CBA256; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23LBN257; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25IBW100; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25IBH103; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25HBN347; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25GBS779; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23KBO199; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23KBN420; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23KBF177; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23FBU813; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23FBP021; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23BBJ851; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23BBJ884; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22LBB596; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22HBR849; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22GBQ034; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22GBQ036; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22EBA005; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22DBU733; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22EBA141; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25EBN366; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25DBT886; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25CBV172; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25ABF384; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24LBM934; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24JBL515; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24HBM973; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24HBG658; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24CBH571; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24CBH400; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24CBH416; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 26BBQ345; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25LBS099; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25JBU372; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25IBT895; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25HBP096; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 26BBP163; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 26BBG647; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25KBR622; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25JBM109; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25FBO089; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25FBG065; Medline Kit Number/SKU DYNJ906037A: UDI/DI each 10193489766202, UDI/DI case 40193489766203, Lot Number: 21BBF015; Medline Kit Number/SKU DYNJ907076D: UDI/DI each 10198459045103, UDI/DI case 40198459045104, Lot Number: 24KBQ125; Medline Kit Number/SKU DYNJ907076D: UDI/DI each 10198459045103, UDI/DI case 40198459045104, Lot Number: 24JBT924; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25HBT028; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25GBG354; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25FBJ135; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25BBD094; Medline Kit Number/SKU DYNJ907446: UDI/DI each 10193489419726, UDI/DI case 40193489419727, Lot Number: 22HBS670; Medline Kit Number/SKU DYNJ907863: UDI/DI each 10193489913187, UDI/DI case 40193489913188, Lot Number: 22EBO903; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 23IBH949; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 23EBS103; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 24ABT534; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 24ABC072; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26CDA182; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26BDA043; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26ADB372; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26ADA380; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25LDA636; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25LDA379; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25JDB762; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25GDA732; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25EDA402; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25BDA442; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25ADB889; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24KDB984; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24KDA219; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24JDC398; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24HDC065; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24GDB441; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24GDA311; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24ADB838; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 23JBD316; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 23DBF809; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24KBU152; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24IBH163; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24HBV805; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24GBC814; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24EBR806; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 26BBP159; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 26ABH446; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25KBN794; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25IBR922; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25EBA894; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25ABL737; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 25BMI762; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 25BMD422; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 24KMG059; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 24KMB410; Medline Kit Number/SKU DYNJ909472C: UDI/DI each 10198459454226, UDI/DI case 40198459454227, Lot Number: 25IMH655; Medline Kit Number/SKU DYNJ909472C: UDI/DI each 10198459454226, UDI/DI case 40198459454227, Lot Number: 25IMC642; Medline Kit Number/SKU DYNJ909926: UDI/DI each 10195327506827, UDI/DI case 40195327506828, Lot Number: 23LBF552; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25FBV690; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25EBK630; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25ABV651; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25ABH588; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 26ABF779; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 25LBR659; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 25LBC067; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 25JBN795; Medline Kit Number/SKU DYNJ910902B: UDI/DI each 10198459387142, UDI/DI case 40198459387143, Lot Number: 25GMB865; Medline Kit Number/SKU DYNJ910987A: UDI/DI each 10198459418532, UDI/DI case 40198459418533, Lot Number: 26AMB048; Medline Kit Number/SKU DYNJ910987A: UDI/DI each 10198459418532, UDI/DI case 40198459418533, Lot Number: 25HMG101; Medline Kit Number/SKU DYNJ911469: UDI/DI each 10198459358753, UDI/DI case 40198459358754, Lot Number: 25JBD450; Medline Kit Number/SKU DYNJ911469: UDI/DI each 10198459358753, UDI/DI case 40198459358754, Lot Number: 25FBS379; Medline Kit Number/SKU DYNJT4005: UDI/DI each 10198459381584, UDI/DI case 40198459381585, Lot Number: 26ABH219; Medline Kit Number/SKU DYNJT4005: UDI/DI each 10198459381584, UDI/DI case 40198459381585, Lot Number: 25GBC350; Medline Kit Number/SKU DYNJT6012: UDI/DI each 10198459531316, UDI/DI case 40198459531317, Lot Number: 25LBN117; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22JDA197; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22HDC165; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22HDB422; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22EDA722; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 21JDA283; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23KDA098; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23IDB014; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23GDA586; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23FDB544; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23DDB294; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23DDB035; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23CDA388; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 22JDB847; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 26ADB209; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25KDB765; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25JDA968; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25IDA858; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25HDA887; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25HDA137; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDB525; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDB387; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDB122; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDA061; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25DDB338; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25CDB735; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25CDA205; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24JDB378; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24IDA577; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24GDA172; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24FDA640; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24BDB046; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24ADA265.
Worldwide distribution.
22116 kits