Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. PLASTICS FREE FLAP, Medline Kit Number/SKU CDS983493N; 2. TLIF PACK, Medline Kit Number/SKU DYNJ02476O; 3. DONOR FREE FLAP PACK-LF, Medline Kit Number/SKU DYNJ0426069N; 4. DONOR FREE FLAP PACK-LF, Medline Kit Number/SKU DYNJ0426069Q; 5. MINOR MB PACK-LF, Medline Kit Number/SKU DYNJ0429303L; 6. ENDOSCOPIC SKULL BASE PACK-LF, Medline Kit Number/SKU DYNJ0753948J; 7. RR-H&N FREEFLAP BLCKWELL PK-LF, Medline Kit Number/SKU DYNJ0785741K; 8. MINIMALLY INVASIVE PACK-LF, Medline Kit Number/SKU DYNJ0843063K; 9. FREE FLAP BASIC SET-UP PK, Medline Kit Number/SKU DYNJ0991320K; 10. FREE FLAP BASIC SET-UP PK, Medline Kit Number/SKU DYNJ0991320M; 11. MLD PACK, Medline Kit Number/SKU DYNJ19521Q; 12. SINUS ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ34623C; 13. PMD PACK SURG CNTR-LF, Medline Kit Number/SKU DYNJ35305G; 14. LARYNGOSCOPY PACK, Medline Kit Number/SKU DYNJ37547D; 15. FUNCTIONAL SINUS ENDOSCOPY, Medline Kit Number/SKU DYNJ40406A; 16. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645C; 17. FREE FLAP B PACK, Medline Kit Number/SKU DYNJ40648G; 18. MINOR ENT PACK, Medline Kit Number/SKU DYNJ50808A; 19. SINUS ENDOSCOPY PACK-LF, Medline Kit Number/SKU DYNJ52062; 20. LAP PACK, Medline Kit Number/SKU DYNJ52116B; 21. LAP PACK, Medline Kit Number/SKU DYNJ52116C; 22. ENDOSCOPY PK, Medline Kit Number/SKU DYNJ54719C; 23. ENDOSCOPY PK, Medline Kit Number/SKU DYNJ54719D; 24. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917I; 25. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917J; 26. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917K; 27. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917L; 28. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917N; 29. LARYNGOSCOPY WEST PACK, Medline Kit Number/SKU DYNJ58324A; 30. CCH DECOMPRESSION PACK, Medline Kit Number/SKU DYNJ67558A; 31. CCH DECOMPRESSION PACK, Medline Kit Number/SKU DYNJ67558B; 32. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213; 33. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213D; 34. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213G; 35. EENT PACK, Medline Kit Number/SKU DYNJ68366B; 36. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447; 37. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447A; 38. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447B; 39. UNIVERSITY DISC PACK, Medline Kit Number/SKU DYNJ69824; 40. MAJOR PACK, Medline Kit Number/SKU DYNJ81325; 41. MAJOR PACK, Medline Kit Number/SKU DYNJ81325A; 42. MAJOR PACK, Medline Kit Number/SKU DYNJ81325B; 43. PLANO DIEP PACK, Medline Kit Number/SKU DYNJ81751; 44. PLANO DIEP PACK, Medline Kit Number/SKU DYNJ81751A; 45. FLAP PACK, Medline Kit Number/SKU DYNJ85359; 46. ENDO PACK, Medline Kit Number/SKU DYNJ86125; 47. ENDO PACK, Medline Kit Number/SKU DYNJ86125A; 48. ENT PACK, Medline Kit Number/SKU DYNJ88448; 49. MAJOR BACK PACK, Medline Kit Number/SKU DYNJ88470; 50. UROLOGY DAVINCI, Medline Kit Number/SKU DYNJ901572O; 51. LARYNGOSCOPY/BRONCH, Medline Kit Number/SKU DYNJ902985F; 52. KIT FREE FLAP MOSES, Medline Kit Number/SKU DYNJ905366B; 53. FLAP, Medline Kit Number/SKU DYNJ907076C; 54. TRANSPHENOIDAL, Medline Kit Number/SKU DYNJ908165; 55. TRANSPHENOIDAL, Medline Kit Number/SKU DYNJ908165A; 56. ENDOSCOPY, Medline Kit Number/SKU DYNJ909512; 57. DIEP FLAP, Medline Kit Number/SKU DYNJ910120A; 58. DIEP BREAST, Medline Kit Number/SKU DYNJ911086; 59. WILLIAMS GENERAL, Medline Kit Number/SKU DYNJ940735F; 60. ROBOTIC LAP PROSTATE PACK CHS, Medline Kit Number/SKU DYNJT3101; 61. CUH FLAP ACCESSORY PACK, Medline Kit Number/SKU DYNJT3243.

Recall: Z-2022-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-2022-2026
Event ID
98664
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 6, 2026
Initiation Date
March 19, 2026
Classification Date
April 30, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. PLASTICS FREE FLAP, Medline Kit Number/SKU CDS983493N; 2. TLIF PACK, Medline Kit Number/SKU DYNJ02476O; 3. DONOR FREE FLAP PACK-LF, Medline Kit Number/SKU DYNJ0426069N; 4. DONOR FREE FLAP PACK-LF, Medline Kit Number/SKU DYNJ0426069Q; 5. MINOR MB PACK-LF, Medline Kit Number/SKU DYNJ0429303L; 6. ENDOSCOPIC SKULL BASE PACK-LF, Medline Kit Number/SKU DYNJ0753948J; 7. RR-H&N FREEFLAP BLCKWELL PK-LF, Medline Kit Number/SKU DYNJ0785741K; 8. MINIMALLY INVASIVE PACK-LF, Medline Kit Number/SKU DYNJ0843063K; 9. FREE FLAP BASIC SET-UP PK, Medline Kit Number/SKU DYNJ0991320K; 10. FREE FLAP BASIC SET-UP PK, Medline Kit Number/SKU DYNJ0991320M; 11. MLD PACK, Medline Kit Number/SKU DYNJ19521Q; 12. SINUS ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ34623C; 13. PMD PACK SURG CNTR-LF, Medline Kit Number/SKU DYNJ35305G; 14. LARYNGOSCOPY PACK, Medline Kit Number/SKU DYNJ37547D; 15. FUNCTIONAL SINUS ENDOSCOPY, Medline Kit Number/SKU DYNJ40406A; 16. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645C; 17. FREE FLAP B PACK, Medline Kit Number/SKU DYNJ40648G; 18. MINOR ENT PACK, Medline Kit Number/SKU DYNJ50808A; 19. SINUS ENDOSCOPY PACK-LF, Medline Kit Number/SKU DYNJ52062; 20. LAP PACK, Medline Kit Number/SKU DYNJ52116B; 21. LAP PACK, Medline Kit Number/SKU DYNJ52116C; 22. ENDOSCOPY PK, Medline Kit Number/SKU DYNJ54719C; 23. ENDOSCOPY PK, Medline Kit Number/SKU DYNJ54719D; 24. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917I; 25. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917J; 26. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917K; 27. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917L; 28. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917N; 29. LARYNGOSCOPY WEST PACK, Medline Kit Number/SKU DYNJ58324A; 30. CCH DECOMPRESSION PACK, Medline Kit Number/SKU DYNJ67558A; 31. CCH DECOMPRESSION PACK, Medline Kit Number/SKU DYNJ67558B; 32. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213; 33. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213D; 34. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213G; 35. EENT PACK, Medline Kit Number/SKU DYNJ68366B; 36. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447; 37. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447A; 38. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447B; 39. UNIVERSITY DISC PACK, Medline Kit Number/SKU DYNJ69824; 40. MAJOR PACK, Medline Kit Number/SKU DYNJ81325; 41. MAJOR PACK, Medline Kit Number/SKU DYNJ81325A; 42. MAJOR PACK, Medline Kit Number/SKU DYNJ81325B; 43. PLANO DIEP PACK, Medline Kit Number/SKU DYNJ81751; 44. PLANO DIEP PACK, Medline Kit Number/SKU DYNJ81751A; 45. FLAP PACK, Medline Kit Number/SKU DYNJ85359; 46. ENDO PACK, Medline Kit Number/SKU DYNJ86125; 47. ENDO PACK, Medline Kit Number/SKU DYNJ86125A; 48. ENT PACK, Medline Kit Number/SKU DYNJ88448; 49. MAJOR BACK PACK, Medline Kit Number/SKU DYNJ88470; 50. UROLOGY DAVINCI, Medline Kit Number/SKU DYNJ901572O; 51. LARYNGOSCOPY/BRONCH, Medline Kit Number/SKU DYNJ902985F; 52. KIT FREE FLAP MOSES, Medline Kit Number/SKU DYNJ905366B; 53. FLAP, Medline Kit Number/SKU DYNJ907076C; 54. TRANSPHENOIDAL, Medline Kit Number/SKU DYNJ908165; 55. TRANSPHENOIDAL, Medline Kit Number/SKU DYNJ908165A; 56. ENDOSCOPY, Medline Kit Number/SKU DYNJ909512; 57. DIEP FLAP, Medline Kit Number/SKU DYNJ910120A; 58. DIEP BREAST, Medline Kit Number/SKU DYNJ911086; 59. WILLIAMS GENERAL, Medline Kit Number/SKU DYNJ940735F; 60. ROBOTIC LAP PROSTATE PACK CHS, Medline Kit Number/SKU DYNJT3101; 61. CUH FLAP ACCESSORY PACK, Medline Kit Number/SKU DYNJT3243.

Reason

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Code Info

Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 25EDA675; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 25CDA718; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 25BDA328; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 24KDB408; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 24KDA607; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23HMH658; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23GMD341; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23EMB793; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23CMG372; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23BMI499; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 22JMF967; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 22EMC648; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 22CMD889; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24EMH897; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24DMF264; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24BMJ904; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24AMH650; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23IME074; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23IMA416; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23EME302; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23DMH071; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23CMH752; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23LME133; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25LMB574; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25IMG534; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25EMJ878; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25DME997; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25CMI768; Medline Kit Number/SKU DYNJ0429303L: UDI/DI each 10195327587260, UDI/DI case 40195327587261, Lot Number: 24HMH265; Medline Kit Number/SKU DYNJ0753948J: UDI/DI each 10193489895896, UDI/DI case 40193489895897, Lot Number: 21KMD078; Medline Kit Number/SKU DYNJ0753948J: UDI/DI each 10193489895896, UDI/DI case 40193489895897, Lot Number: 21HME198; Medline Kit Number/SKU DYNJ0753948J: UDI/DI each 10193489895896, UDI/DI case 40193489895897, Lot Number: 21FMD014; Medline Kit Number/SKU DYNJ0785741K: UDI/DI each 10193489603576, UDI/DI case 40193489603577, Lot Number: 21CKA430; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23JDA179; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23HDC666; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23FDB421; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23EDA191; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23CDC020; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23BDA963; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23ADB394; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23LDA280; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21ILA211; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21HLA920; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21HLA656; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21GLA243; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24IMH942; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24IMB140; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24GMB047; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24EMG280; Medline Kit Number/SKU DYNJ19521Q: UDI/DI each 10195327318079, UDI/DI case 40195327318070, Lot Number: 23BMB103; Medline Kit Number/SKU DYNJ19521Q: UDI/DI each 10195327318079, UDI/DI case 40195327318070, Lot Number: 25BMG556; Medline Kit Number/SKU DYNJ34623C: UDI/DI each 10889942745807, UDI/DI case 40889942745808, Lot Number: 23CBT880; Medline Kit Number/SKU DYNJ34623C: UDI/DI each 10889942745807, UDI/DI case 40889942745808, Lot Number: 22KBW690; Medline Kit Number/SKU DYNJ34623C: UDI/DI each 10889942745807, UDI/DI case 40889942745808, Lot Number: 21LBG909; Medline Kit Number/SKU DYNJ35305G: UDI/DI each 10195327070328, UDI/DI case 40195327070329, Lot Number: 23AMB303; Medline Kit Number/SKU DYNJ37547D: UDI/DI each 10889942945993, UDI/DI case 40889942945994, Lot Number: 21DDB912; Medline Kit Number/SKU DYNJ37547D: UDI/DI each 10889942945993, UDI/DI case 40889942945994, Lot Number: 21CDB910; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23JBY647; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23HBX192; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23GBC541; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23DBD185; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23CBP552; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23ABG640; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22JBW980; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22HBY401; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22GBG635; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22EBL969; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22CBX216; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22ABL872; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21KBK659; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21IBR941; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21FBG655; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21DBG029; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21CBJ871; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 26CBE134; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 26ABM247; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25KBF571; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25IBS828; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25HBF128; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25GBD444; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25EBN746; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25DBU598; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25BBP703; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24LBL146; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24HBR568; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24GBW861; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24FBL120; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24EBD749; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23LBI152; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23JMA313; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23HMH325; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23GMA570; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23EMI670; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23DMC678; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23CMB630; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22LMF935; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22HMF705; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22FMD171; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22FMD415; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22EMB584; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22DMF559; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 23DMH298; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22KMH672; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22GMD348; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22FMD240; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22EME295; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22DMG577; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 23HBR726; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 23FBG834; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 23CBL006; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22IBL408; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22HBY173; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22EBF344; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22ABB710; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21LBR187; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21KBE554; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21HBQ602; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21HBQ631; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21GBJ642; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21CBS316; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21BBN842; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 25EBR511; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 25BBU313; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24IBF115; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24GBX974; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24FBR318; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24ABP408; Medline Kit Number/SKU DYNJ52062: UDI/DI each 10889942235742, UDI/DI case 40889942235743, Lot Number: 21LBU391; Medline Kit Number/SKU DYNJ52062: UDI/DI each 10889942235742, UDI/DI case 40889942235743, Lot Number: 21VBD187; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 23JMB108; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 23EMJ664; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 23CMC992; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 22LMF460; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 22FMD641; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 22ABG291; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21KBF791; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21HBT589; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21DBJ867; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21CBA789; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 24FMF952; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 24BMJ713; Medline Kit Number/SKU DYNJ52116C: UDI/DI each 10198459270352, UDI/DI case 40198459270353, Lot Number: 25LMD153; Medline Kit Number/SKU DYNJ52116C: UDI/DI each 10198459270352, UDI/DI case 40198459270353, Lot Number: 25KMH925; Medline Kit Number/SKU DYNJ52116C: UDI/DI each 10198459270352, UDI/DI case 40198459270353, Lot Number: 25BMD068; Medline Kit Number/SKU DYNJ54719C: UDI/DI each 10193489433449, UDI/DI case 40193489433440, Lot Number: 21CBH091; Medline Kit Number/SKU DYNJ54719C: UDI/DI each 10193489433449, UDI/DI case 40193489433440, Lot Number: 21ABM862; Medline Kit Number/SKU DYNJ54719D: UDI/DI each 10193489885293, UDI/DI case 40193489885294, Lot Number: 21HBE730; Medline Kit Number/SKU DYNJ54719D: UDI/DI each 10193489885293, UDI/DI case 40193489885294, Lot Number: 21FBO645; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21GDB034; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21FDC469; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21EDC231; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21ADA830; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22KDA276; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22JDA493; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22FDB151; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22EDB139; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22DDA699; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22BDA826; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 21KDB492; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 21IDB626; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23EDC310; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23DDA529; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23BDC220; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23ADB232; Medline Kit Number/SKU DYNJ56917L: UDI/DI each 10195327466855, UDI/DI case 40195327466856, Lot Number: 23KDB950; Medline Kit Number/SKU DYNJ56917L: UDI/DI each 10195327466855, UDI/DI case 40195327466856, Lot Number: 23JDB895; Medline Kit Number/SKU DYNJ56917L: UDI/DI each 10195327466855, UDI/DI case 40195327466856, Lot Number: 24BDB140; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25IMB406; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25EMA512; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25DMG834; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25CMC310; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25BMC774; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25BMC034; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 24JDB128; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 23FBS890; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 23DBO019; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 22IBG087; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 22BBZ926; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 21CBO062; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 21CBT626; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 21ABA621; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 25FBB275; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24HBQ403; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24HBC985; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24FBB168; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24EBK104; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24CBG181; Medline Kit Number/SKU DYNJ67558A: UDI/DI each 10193489917437, UDI/DI case 40193489917438, Lot Number: 21FBF038; Medline Kit Number/SKU DYNJ67558B: UDI/DI each 10195327099053, UDI/DI case 40195327099054, Lot Number: 22OBG478; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22EMA482; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22CMI811; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22CMG430; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22AMB156; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21KMB789; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21IMG069; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21IMA320; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21GMF680; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21GMC469; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21FMB885; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21EME095; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21DMB893; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21BMC546; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 25FMB744; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 25DMG846; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 25BMB216; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 24KMI278; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 24JMG883; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 24JMC380; Medline Kit Number/SKU DYNJ68213G: UDI/DI each 10198459648175, UDI/DI case 40198459648176, Lot Number: 26AMA296; Medline Kit Number/SKU DYNJ68366B: UDI/DI each 10195327666330, UDI/DI case 40195327666331, Lot Number: 24EMB363; Medline Kit Number/SKU DYNJ68447: UDI/DI each 10193489447095, UDI/DI case 40193489447096, Lot Number: 21CBL485; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21VBB288; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21IBV467; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21IBG628; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21GBN769; Medline Kit Number/SKU DYNJ68447B: UDI/DI each 10195327034160, UDI/DI case 40195327034161, Lot Number: 22NBF961; Medline Kit Number/SKU DYNJ68447B: UDI/DI each 10195327034160, UDI/DI case 40195327034161, Lot Number: 22ABV162; Medline Kit Number/SKU DYNJ68447B: UDI/DI each 10195327034160, UDI/DI case 40195327034161, Lot Number: 21LBR924; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23JBU091; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23JBC614; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23EBR266; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23EBB687; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23CBV202; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23BBP560; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23ABL869; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23ABI200; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22KBJ128; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22IBU121; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22HBY433; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22GBR086; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22GBQ113; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22FBP609; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22BBU834; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25FBB994; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25DBQ190; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25CBN006; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25ABQ389; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24KBO095; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24JBQ827; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24IBH741; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24HBO248; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24GBK688; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24GBE944; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24FBQ277; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24EBO557; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24EBB354; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24BBU070; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23LBT175; Medline Kit Number/SKU DYNJ81325: UDI/DI each 10195327111014, UDI/DI case 40195327111015, Lot Number: 22JBC490; Medline Kit Number/SKU DYNJ81325: UDI/DI each 10195327111014, UDI/DI case 40195327111015, Lot Number: 22GBW885; Medline Kit Number/SKU DYNJ81325: UDI/DI each 10195327111014, UDI/DI case 40195327111015, Lot Number: 22EBA478; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23KBD697; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23JBC513; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23IBD507; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23FBS833; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23DBS310; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23CBW627; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23BBO729; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 22KBT451; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 22KBD178; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24KBA572; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24JBH042; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24IBT346; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24FBJ124; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24EBN022; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24DBL535; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24BBS543; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24ABN491; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23LBF176; Medline Kit Number/SKU DYNJ81325B: UDI/DI each 10198459217029, UDI/DI case 40198459217020, Lot Number: 25FBD987; Medline Kit Number/SKU DYNJ81325B: UDI/DI each 10198459217029, UDI/DI case 40198459217020, Lot Number: 25EBD835; Medline Kit Number/SKU DYNJ81325B: UDI/DI each 10198459217029, UDI/DI case 40198459217020, Lot Number: 25BBE525; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 23ABA516; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 22JBE441; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 22JBD573; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 22GBX010; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23JBW345; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23JBG162; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23FBA908; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23EBW504; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 26ABU708; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25LBM684; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25KBQ995; Medline Kit Number/SKU DYNJ81751A: UDI/DI e

More Code Info

ach 10195327345112, UDI/DI case 40195327345113, Lot Number: 25IBV639; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25IBU643; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25HBT909; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25FBG513; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25CBQ697; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25ABE335; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24KBE398; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24JBD820; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24EBU640; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24DBT114; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24ABO861; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23LBJ179; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 23KBM252; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 23HBX929; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 26CBD685; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 26ABE085; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25KBN912; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25JBI743; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25GBP180; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25FBL455; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25FBA096; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25DBR251; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25BBT099; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24LBE877; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24KBL906; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24IBF438; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24HBQ248; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24GBO711; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24FBS345; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24DBA574; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24ABN402; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 23LBK140; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25GBA828; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25EBF285; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25CBU143; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25ABG027; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24KBN578; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24HBP866; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24FBJ095; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24DBG675; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24CBL699; Medline Kit Number/SKU DYNJ86125A: UDI/DI each 10198459473852, UDI/DI case 40198459473853, Lot Number: 26AMC689; Medline Kit Number/SKU DYNJ86125A: UDI/DI each 10198459473852, UDI/DI case 40198459473853, Lot Number: 25LMD206; Medline Kit Number/SKU DYNJ86125A: UDI/DI each 10198459473852, UDI/DI case 40198459473853, Lot Number: 25JME672; Medline Kit Number/SKU DYNJ88448: UDI/DI each 10198459161322, UDI/DI case 40198459161323, Lot Number: 25JBC138; Medline Kit Number/SKU DYNJ88448: UDI/DI each 10198459161322, UDI/DI case 40198459161323, Lot Number: 24JBP409; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25LMJ318; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25KMF373; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25JMC356; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25IMA237; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 24KMA470; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22LMJ185; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22KMH569; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22JMH343; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22JME988; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22JMA764; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22IMG572; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22GBD794; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22EBB404; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22EBC306; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22DBI454; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22CBQ232; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22ABG615; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21LBN116; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBZ349; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBM717; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBH386; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBE452; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21GBG956; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21EBI236; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21DBP367; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21BBC494; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 23KMB541; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 23IMI202; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22JMG646; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22EMF849; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22EMD560; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22BME333; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21JMA467; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21HME157; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21FMB959; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21DKA711; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21DKA257; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21AKA915; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 23LMG587; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23EBM739; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23DBM835; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23CBE807; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22LBO631; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22HBM673; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22GBE117; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22GBA660; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22DBR510; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 25DBG803; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 25ABJ890; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 25ABF688; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24JBP878; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24HBB373; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24GBH830; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24EBN295; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24EBD925; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24ABT424; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23LBT823; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23IBH134; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23EBW674; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23DBM339; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23ABR108; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23ABM208; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 22HBS616; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 22IBC655; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 22GBG134; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 21KBV892; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 21JBS139; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 24DBS653; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 24CBQ078; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 24ABS092; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23KBH043; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23KBH076; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23IBO315; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23FBA102; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBL712; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBG445; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBF440; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBD054; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22JBI497; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22FLA058; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22CLA782; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22CLA236; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23LBG978; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25LBE981; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25JBS925; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25IBS527; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25IBC316; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25EBR807; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25DBE206; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25ABF294; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24IBV349; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24GBQ643; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24EBT890; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24EBH261; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24EBC018; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24DBG691; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24ABD445; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 23FBA333; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 23EBF184; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 26CBF295; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25IBT900; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25IBD206; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25GBR222; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25DBS691; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25ABQ137; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25ABJ070; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 24IBP825; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 24CBT484; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 23LBD654; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25LMG501; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25JMG389; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25IMJ182; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25HMC859; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25GMD539; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25EME174; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25DME221; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25BME721; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 24LMB897; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 24JMI715; Medline Kit Number/SKU DYNJ911086: UDI/DI each 10198459248757, UDI/DI case 40198459248758, Lot Number: 25EBI830; Medline Kit Number/SKU DYNJ911086: UDI/DI each 10198459248757, UDI/DI case 40198459248758, Lot Number: 25CBO097; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23DBF803; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23CBN885; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23BBT129; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23ABC272; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22LBC489; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22JBM643; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22IBH997; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22HBJ672; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22GBH649; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22FBT217; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22DBK649; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 26BBA753; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 26ABA848; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25JBE538; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25HBS777; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25FBO606; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25EBR590; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25EBK899A; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25JBR004; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25JBB805; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25HBL978; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25EBK899.

Distribution

Worldwide distribution.

Quantity

21288 kits