FDA Enforcement Class II Terminated

11cm QD Angle Attachment Product Usage: Pneumatic system

Recall: Z-2020-2017 · Reported May 10, 2017

Enforcement

Recall Number
Z-2020-2017
Event ID
76059
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
The Anspach Effort, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 10, 2017
Initiation Date
October 6, 2016
Classification Date
May 4, 2017
Termination Date
July 26, 2019
Address
4500 Riverside Dr, Palm Beach Gardens, FL, 33410-4235, United States

Description

11cm QD Angle Attachment Product Usage: Pneumatic system

Reason

Supplied Directions for Use (DFU) did not contain a recommended service interval.

Code Info

All serial numbers shipped to customers prior to 09/27/2016; DFU 18-0048; Part No. QD11.

Distribution

Worldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan

Quantity

455