FDA Enforcement Class II Terminated

AXIOM Artis dTA, Model Number 7008605

Recall: Z-2019-2019 · Reported July 31, 2019

Enforcement

Recall Number
Z-2019-2019
Event ID
83203
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2019
Initiation Date
June 12, 2019
Classification Date
July 22, 2019
Termination Date
September 18, 2020
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

AXIOM Artis dTA, Model Number 7008605

Reason

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Code Info

Serial Numbers: 55163 55713 55304 55435 55567 55407 55412 55310 55470 55164 55187 55206 55091 55059 55204 55377 55307 55171 55368 55210 55133 55135 55477 55250 55318 55196 55230 55277 55423 55320 55387 55576 55438 55503 55046 55308 55139 55039 55270 55188 55314 55381 55550 55365 55574 55415 55109 55337 55322 55272 55157 55234 55548 55282 55575 55211 55410 55150 55079 55262 55259 55531 55388 55275 55276 55254 55569 55177 55269 55292 55041 55404 55425 55183 55441 55125 55369 55437 55136 55356 55191 55103 55400 55483 55083 55526 55088 55243 55056 55408 55715 55233 55416 55256 55074 55212 55153 55325 55312 55273 55213 55401 55170 55284 55469 55067 55048 55216 55241 55708 55202 55461 55154 55167 55342 55468 55219 55553 55217 55386 55396 55513 55279 55414 55347 55042 55294 55160 55286 55036 55362 55436 55375 55131 55062 55107 55411 55052 55096 55156 55289 55374 55705 55017 55223 55149 55141 55089 55305 55078 55227 55040 55200 55186 55184 55232 55358 55222 55313 55085 55090 55324 55045 55382 55350 55353 55405 55043 55439 55473 55070 55545 55434 55118 55455 55263 55274 55238

Distribution

US Nationwide Distribution

Quantity

178