FDA Enforcement Class II Terminated

Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other devices via Flexport interfaces. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Spacelabs Healthcare Xprezzon Bedside Monitor may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitor allows network based applications to open windows and display information the Xprezzon and other networked monitors. The Spacelabs Healthcare Xprezzon Bedside Monitor is designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Xprezzon Bedside Monitors is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

Recall: Z-2019-2012 · Reported July 25, 2012

Enforcement

Recall Number
Z-2019-2012
Event ID
62437
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Spacelabs Healthcare, Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 25, 2012
Initiation Date
June 12, 2012
Classification Date
July 16, 2012
Termination Date
November 28, 2012
Address
5150 220th Ave Se, Issaquah, WA, 98029-6834, United States

Description

Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other devices via Flexport interfaces. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Spacelabs Healthcare Xprezzon Bedside Monitor may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitor allows network based applications to open windows and display information the Xprezzon and other networked monitors. The Spacelabs Healthcare Xprezzon Bedside Monitor is designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Xprezzon Bedside Monitors is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

Reason

There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.

Code Info

040-1632-00; 1393-000013; 1393-000039; 1393-000040; 1393-000041; 1393-000341; 1393-000382; 1393-000383; 1393-000384; 1393-000385; 1393-000386; 1393-000387; 1393-000388; 1393-000808; 1393-000831; 1393-001374; 1393-001375; 1393-001453; 1393-001990; 1393-001991; and 1393-001992.

Distribution

Worldwide Distribution - USA including Florida, Iowa, Oklahoma, South Caroline and the countries of India, Germany and Panama.

Quantity

21 units