FDA Enforcement Class II Ongoing

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PACK, Medline Kit Number/SKU DYNJ62635B.

Recall: Z-2018-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-2018-2026
Event ID
98664
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 6, 2026
Initiation Date
March 19, 2026
Classification Date
April 30, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PACK, Medline Kit Number/SKU DYNJ62635B.

Reason

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Code Info

Medline Kit Number/SKU DYNJ32854B: UDI/DI each 10198459413414, UDI/DI case 40198459413415, Lot Number : 25GBU656; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25GMD214; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25FMC376; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25DMJ030; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25BME902; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25AMJ032; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24IMH612; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24EME486; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24DMD315; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 26BMD607; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 25JMD196.

Distribution

Worldwide distribution.

Quantity