FDA Enforcement Class II Ongoing

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Recall: Z-2005-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-2005-2026
Event ID
98680
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Healthcare Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2026
Initiation Date
April 2, 2026
Classification Date
April 29, 2026
Address
2 Edgewater Dr, Norwood, MA, 02062-4637, United States

Description

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Reason

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

Code Info

Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);

Distribution

Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.

Quantity

71725 (38025 US, 33700 OUS)