PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Recall: Z-2004-2023 · Reported July 5, 2023

Enforcement

Recall Number
Z-2004-2023
Event ID
92250
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Permobil
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 5, 2023
Initiation Date
May 10, 2023
Classification Date
June 26, 2023
Address
300 Duke Dr, N/A, Lebanon, TN, 37090-8115, United States

Description

PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application

Reason

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

Code Info

All units with software versions 1.0.00 to 1.0.04

Distribution

Worldwide distribution.

Quantity

18,905 units