FDA Enforcement Class II Ongoing

i-STAT G3+ cartridge; List Number: 03P78-26;

Recall: Z-2001-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-2001-2026
Event ID
98635
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Point Of Care Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2026
Initiation Date
April 1, 2026
Classification Date
April 29, 2026
Address
400 College Rd E, Princeton, NJ, 08540-6607, United States

Description

i-STAT G3+ cartridge; List Number: 03P78-26;

Reason

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

Code Info

List Number: 03P78-26; UDI-DI: 00054749005673; Lot number: N25221;

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.

Quantity

40275 units (US only)