65597201131 Femur CR POR FEM SURF HDN HATCP AML Rx, Sterile; 65597201132 Femur CR POR FEM SURF HDN HATCP AMR Rx, Sterile; 65597201231 Femur CR POR FEM SURF HDN HATCP BML Rx, Sterile; 65597201232 Femur CR POR FEM SURF HDN HATCP BMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Enforcement
- Recall Number
- Z-2001-2013
- Event ID
- 65413
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 28, 2013
- Initiation Date
- June 7, 2013
- Classification Date
- August 22, 2013
- Termination Date
- December 15, 2014
- Address
- 345 E Main St, N/A, Warsaw, IN, 46580-2746, United States
Description
65597201131 Femur CR POR FEM SURF HDN HATCP AML Rx, Sterile; 65597201132 Femur CR POR FEM SURF HDN HATCP AMR Rx, Sterile; 65597201231 Femur CR POR FEM SURF HDN HATCP BML Rx, Sterile; 65597201232 Femur CR POR FEM SURF HDN HATCP BMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
all codes
Worldwide Distribution-USA (nationwide) and the country of Japan.
192,355