i-STAT EG7+ cartridge; List Number: 03P76-25;
Enforcement
- Recall Number
- Z-1999-2026
- Event ID
- 98635
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Abbott Point Of Care Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2026
- Initiation Date
- April 1, 2026
- Classification Date
- April 29, 2026
- Address
- 400 College Rd E, Princeton, NJ, 08540-6607, United States
Description
i-STAT EG7+ cartridge; List Number: 03P76-25;
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
List Number: 03P76-25; UDI-DI: 00054749000043; Lot numbers: N25206, N25207, N25209, N25213, N25218, N25219, N25233;
Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.
369650 units (150875 US, 218775 OUS)