Heraeus, PALAMIX uno. Material Number: 66057893.
Enforcement
- Recall Number
- Z-1997-2026
- Event ID
- 98608
- Classification
- Class III
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Heraeus Medical GmbH (Dental Division)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2026
- Initiation Date
- March 25, 2026
- Classification Date
- April 29, 2026
- Address
- Philipp-Reis-Str. 8, Wehrheim, N/A, Germany
Description
Heraeus, PALAMIX uno. Material Number: 66057893.
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
51,771 units