HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
Recall: Z-1995-2014
·
Reported July 16, 2014
Enforcement
- Recall Number
- Z-1995-2014
- Event ID
- 68580
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cenorin, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 16, 2014
- Initiation Date
- January 24, 2013
- Classification Date
- July 6, 2014
- Termination Date
- February 10, 2015
- Address
- 6324 S 199th Pl Ste 107, Kent, WA, 98032-2135, United States
Description
HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
Reason
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
Code Info
610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.
Distribution
Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
Quantity
25 HLD systems