FDA Enforcement Class II Terminated

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Recall: Z-1995-2014 · Reported July 16, 2014

Enforcement

Recall Number
Z-1995-2014
Event ID
68580
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cenorin, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 16, 2014
Initiation Date
January 24, 2013
Classification Date
July 6, 2014
Termination Date
February 10, 2015
Address
6324 S 199th Pl Ste 107, Kent, WA, 98032-2135, United States

Description

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Reason

The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Code Info

610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.

Distribution

Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.

Quantity

25 HLD systems