Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.
Enforcement
- Recall Number
- Z-1994-2013
- Event ID
- 65866
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Getinge USA Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 28, 2013
- Initiation Date
- August 2, 2013
- Classification Date
- August 21, 2013
- Termination Date
- February 23, 2017
- Address
- 1777 E Henrietta Rd, N/A, Rochester, NY, 14623-3133, United States
Description
Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.
AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters.
LOT SR-430 (Expiration 02/06/2014)
Worldwide Distribution - USA and the country of Canada.
Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs)