FDA Enforcement
Class II
Terminated
Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041
Recall: Z-1992-2016
·
Reported June 22, 2016
Enforcement
- Recall Number
- Z-1992-2016
- Event ID
- 73692
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Med Tec Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 22, 2016
- Initiation Date
- March 31, 2016
- Classification Date
- June 15, 2016
- Termination Date
- February 3, 2017
- Address
- PO Box 320, 1401 8th St SE, Orange City, IA, 51041-7463, United States
Description
Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041
Reason
When using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.
Code Info
Versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1)
Distribution
Worldwide Distribution - US Distribution including the countries of Switzerland, Germany..
Quantity
9