FDA Enforcement Class II Ongoing

ICU Medical Plum Solos, Item number 400011001; infusion pump

Recall: Z-1991-2025 · Reported June 25, 2025

Enforcement

Recall Number
Z-1991-2025
Event ID
96777
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Visit
Report Date
June 25, 2025
Initiation Date
April 22, 2025
Classification Date
June 18, 2025
Address
600 N Field Dr, N/A, Lake Forest, IL, 60045-4835, United States

Description

ICU Medical Plum Solos, Item number 400011001; infusion pump

Reason

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Code Info

UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173, 60000174, 60000176, 60000177, 60000178, 60000179, 60000180, 60000183, 60000187, 60000189, 60000190, 60000191, 60000192, 60000198, 60000199, 60000201, 60000203, 60000204, 60000205, 60000206, 60000207, 60000208, 60000209, 60000210, 60000212, 60000213, 60000214, 60000215, 60000216, 60000217, 60000223, 60000224, 60000225, 60000226, 60000227, 60000228, 60000229, 60000230, 60000234, 60000236, 60000237, 60000238, 60000239, 60000240, 60000241, 60000242, 60000244, 60000245, 60000246, 60000247, 60000249, 60000251, 60000252, 60000255, 60000257, 60000259, 60000263, 60000264, 60000265, 60000266, 60000268, 60000269, 60000271, 60000273, 60000274, 60000275, 60000276, 60000277, 60000279, 60000280, 60000281, 60000282, 60000283, 60000284, 60000285, 60000286, 60000288, 60000291, 60000292, 60000295, 60000296, 60000297, 60000298, 60000300, 60000155, 60000193, 60000197, 60000211, 60000243, 60000270, 60000272, 60000123, 60000168, 60000175, 60000195, 60000221, 60000256, 60000294.

Distribution

Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

Quantity

214 units