FDA Enforcement Class II Ongoing

Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10. Mobile X-ray System.

Recall: Z-1991-2015 · Reported August 12, 2015

Enforcement

Recall Number
Z-1991-2015
Event ID
71649
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Scientific Solutions Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 12, 2015
Initiation Date
June 9, 2015
Classification Date
July 31, 2015
Address
48 Woerd Ave, N/A, Waltham, MA, 02453-3824, United States

Description

Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10. Mobile X-ray System.

Reason

Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).

Code Info

Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10.

Distribution

Nationwide Distribution.

Quantity

1047