FDA Enforcement Class II Terminated

Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Recall: Z-1989-2020 · Reported May 20, 2020

Enforcement

Recall Number
Z-1989-2020
Event ID
85629
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Covidien Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 20, 2020
Initiation Date
April 30, 2020
Classification Date
May 14, 2020
Termination Date
December 14, 2023
Address
15 Hampshire St, N/A, Mansfield, MA, 02048-1113, United States

Description

Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Reason

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Code Info

Lot Numbers: 1909308664, 1909320964 1933802964

Distribution

US Nationwide distribution.

Quantity

8162 units