FDA Enforcement
Class II
Ongoing
Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.A
Recall: Z-1987-2023
·
Reported June 28, 2023
Enforcement
- Recall Number
- Z-1987-2023
- Event ID
- 92325
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Olympus Corporation of the Americas
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 28, 2023
- Initiation Date
- May 4, 2023
- Classification Date
- June 21, 2023
- Address
- 3500 Corporate Pkwy, Center Valley, PA, 18034-8229, United States
Description
Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.A
Reason
Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.
Code Info
UDI-DI: 04953170385919 Lot Numbers: 24K, 27K, 29K, 2XK, 2YK
Distribution
US Nationwide distribution.
Quantity
340 pieces