Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
Enforcement
- Recall Number
- Z-1986-2021
- Event ID
- 87974
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America Llc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 7, 2021
- Initiation Date
- May 1, 2021
- Classification Date
- June 30, 2021
- Address
- 222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States
Description
Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
Model/ Serial Number/UDI: N/A 10016 (01)00884838101111(21)10016 10018 (01)00884838101111(21)10018
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.
1,301 Monitor Bracket Assemblies total