FDA Enforcement Class II Ongoing

BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use

Recall: Z-1983-2026 · Reported May 6, 2026

Enforcement

Recall Number
Z-1983-2026
Event ID
98657
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton Dickinson & Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2026
Initiation Date
April 7, 2026
Classification Date
April 29, 2026
Address
7 Loveton Cir, Sparks, MD, 21152-9212, United States

Description

BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use

Reason

BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.

Code Info

Catalog Number: 260685; UDI-DI: 30382902606858; Lot Number: 5329863.

Distribution

Worldwide distribution - US Nationwide and the country of Hong Kong.

Quantity

2,200 units