MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel
Enforcement
- Recall Number
- Z-1982-2023
- Event ID
- 92327
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 28, 2023
- Initiation Date
- May 3, 2023
- Classification Date
- June 21, 2023
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel
Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.
a) REF MNK0005: GTIN 30884389161566, Lot Numbers: LMPTH0189, LMPTH0191, LMPTH0193, LMPTH0195, LMPTH0197, LMPTH0199, LMPTH0201, LMPTH0203, LMPTH0205, LMPTH0207, LMPTH0211, LMPTH0213; b) REF MNK0015: GTIN 40884389154558, Lot Numbers: LMPTH0188, LMPTH0190, LMPTH0192, LMPTH0196, LMPTH0198, LMPTH0200, LMPTH0202, LMPTH0204, LMPTH0206, LMPTH0210, LMPTH0210B, LMPTH0212
Worldwide distribution - US Nationwide and the countries of SINGAPORE, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, DUBAI, CHINA, PANAMA, CANADA, SAUDI ARABIA, BERMUDA, COLOMBIA, MAURITIUS, KUWAIT.
851,328 units