FDA Enforcement
Class II
Ongoing
V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).
Recall: Z-1980-2023
·
Reported June 28, 2023
Enforcement
- Recall Number
- Z-1980-2023
- Event ID
- 92355
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- SenTec AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 28, 2023
- Initiation Date
- May 2, 2023
- Classification Date
- June 20, 2023
- Address
- Ringstrasse 39, Therwil, N/A, Switzerland
Description
V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).
Reason
The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.
Code Info
Serial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.
Distribution
US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.
Quantity
11 in the U.S.